![Fluke NetDAQ 2640A Service Manual Download Page 183](http://html1.mh-extra.com/html/fluke/netdaq-2640a/netdaq-2640a_service-manual_2310217183.webp)
Diagnostic Testing and Troubleshooting
Troubleshooting the Instrument
5
5-19
Command: MEAS?
Description: Takes a measurement on channel one and returns the measured value. This
command is new to the instrument.
Parameters: None
Response: <measurement> = IEEE-488.2 NR3 ASCII representation of a floating point
value: leading sign;
5 or 6 digits with embedded decimal point for 2645A and 2640A respectively,
padded
with leading and trailing zeroes if necessary;
“E”;
a signed exponent of -3, +0, +3, or +6.
For e115.67E+0.
An overload condition is indicated by a value of +001.00E+9 or -001.00E+9 in the
2645A, +001.000E+9 or -001.000E+9 for the 2640A.
Restrictions: Not allowed in calibration mode. If channel one is not configured, this
command generates an execution error and does not return a value.
Notes: While processing this query, the instrument does not return a response or a
prompt until the measurement has been taken.
The instrument generates a device dependent error if the measurement does not arrive
after 30 seconds.
Open thermocouples return a +999.99[9]E+9.
Troubleshooting the Instrument
5-19.
The following paragraphs describe an organized method of instrument troubleshooting.
The overall approach is to start with power-up self-test error codes and then proceed to
more and more detailed procedures. Begin troubleshooting with “General
Troubleshooting.” These procedures locate about 90% of the instrument faults. The
remaining faults require sleuthing that is beyond the scope of this manual. For these
cases, review Chapter 2 “Theory of Operation” and approach the difficulty in a logical
manner adapting the troubleshooting procedures, as required. As a last resort, contact the
factory or Fluke Service Center (see Chapter 6) for assistance.
General Troubleshooting
5-20.
General troubleshooting uses the instrument response to self-test as a clue to the fault
location. If the instrument completes self-test and displays an error code on the front
panel, refer to Table 5-10. If the instrument appears dead and will not even self test, refer
to Table 5-11. If the instrument passes self test but is not operating correctly, then refer
to “Troubleshooting the A3 A/D Converter PCA” for analog problems, and
“Troubleshooting the A1 Main PCA” for digital problems. For assembly and
disassembly procedures, see Chapter 3, “General Maintenance.”
Summary of Contents for NetDAQ 2640A
Page 14: ...NetDAQ Service Manual x...
Page 46: ...NetDAQ Service Manual 2 4...
Page 106: ...NetDAQ Service Manual 2 64...
Page 108: ...NetDAQ Service Manual 3 2...
Page 164: ...NetDAQ Service Manual 4 42...
Page 206: ...NetDAQ Service Manual 6 2...
Page 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Page 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Page 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Page 234: ...NetDAQ Service Manual 7 2...
Page 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Page 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Page 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Page 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Page 260: ...2640A 2645A Service Manual 7 28...