![Fluke NetDAQ 2640A Service Manual Download Page 150](http://html1.mh-extra.com/html/fluke/netdaq-2640a/netdaq-2640a_service-manual_2310217150.webp)
NetDAQ
Service Manual
4-28
There are two connection methods used between the leads from the instrument Universal
Input Module and the 5700A Calibrator. The volts dc, volts ac, and resistance functions
use the two-wire connection shown in Figure 4-7. The resistance function uses the four-
wire connection shown in Figure 4-8.
Ending Calibration
4-33.
When you are using the manual method of calibration and have completed the
calibration of the desired functions (dc volts, ac volts, ohms, and frequency), press the
front panel CAL Enable button again to clear CAL from the primary display. When
using the calibration features of NetDAQ Logger for Windows, calibration is ended
automatically when you exit the calibration mode.
RS-232 Instrument Configuration Parameters
4-34.
When CAL is shown in the primary display indicating you are in the calibration mode,
the instrument configuration parameters are set as listed in Table 4-2.
Table 4-2. RS-232 Instrument Configuration for Calibration Procedures
Parameter
Setting
Channel Configuration
Unique to calibration step
Reading Rate
Slow
Trigger Type
All disabled
Intervals
All disabled
Drift Correction
Enabled
External Trigger Output
Disabled
Temperature Units
Celsius
Queue Overflow Mode
Discard old
Totalizer Debounce
Disabled
Autodisable Scanning
On
Calibration Procedure (Automatic)
4-35.
Automatic calibration uses Fluke MET/CAL software. All procedures are provided in
the MET/CAL Users Manual. These procedures assume you have completed the
“Preparing for Calibration” procedure above.
Calibration Procedure (Semiautomatic)
4-36.
Semiautomatic calibration uses the calibration feature built into the NetDAQ Logger for
Windows software. This procedure assumes you have completed the “Preparing for
Calibration” procedure above.
Complete the following procedure to calibrate an instrument using NetDAQ Logger for
Windows software. This procedure integrates the four calibration cycles: VDC, VAC,
ohms, and frequency. Complete only the cycle or cycles of interest.
1.
If you have not already done so, complete the “Preparing for Calibration” procedure
earlier in this chapter. Note the instrument baud rate and host computer COM port
used for interconnection.
Summary of Contents for NetDAQ 2640A
Page 14: ...NetDAQ Service Manual x...
Page 46: ...NetDAQ Service Manual 2 4...
Page 106: ...NetDAQ Service Manual 2 64...
Page 108: ...NetDAQ Service Manual 3 2...
Page 164: ...NetDAQ Service Manual 4 42...
Page 206: ...NetDAQ Service Manual 6 2...
Page 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Page 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Page 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Page 234: ...NetDAQ Service Manual 7 2...
Page 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Page 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Page 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Page 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Page 260: ...2640A 2645A Service Manual 7 28...