Fluke I430FLEX Instruction Sheet Download Page 1

LIMITED WARRANTY & LIMITATION OF LIABILITY

LIMITED WARRANTY & LIMITATION OF LIABILITY

 

 
Each Fluke product is warranted to be free from defects in material and 
workmanship under normal use and service. The warranty period is one year 
and begins on the date of shipment. Parts, product repairs and services are 
warranted for 90 days. This warranty extends only to the original buyer or 
end-user customer of a Fluke authorized reseller, and does not apply to 
fuses, disposable batteries or to any product which, in Fluke's opinion, has 
been misused, altered, neglected or damaged by accident or abnormal 
conditions of operation or handling. Fluke warrants that software will operate 
substantially in accordance with its functional specifications for 90 days and 
that it has been properly recorded on non-defective media. Fluke does not 
warrant that software will be error free or operate without interruption.  
 
Fluke authorized resellers shall extend this warranty on new and unused 
products to end-user customers only but have no authority to extend a greater or 
different warranty on behalf of Fluke. Warranty support is available if product is 
purchased through a Fluke authorized sales outlet or Buyer has paid the 
applicable international price. Fluke reserves the right to invoice Buyer for 
importation costs of repair/replacement parts when product purchased in one 
country is submitted for repair in another country. 
 
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase 
price, free of charge repair, or replacement of a defective product which is 
returned to a Fluke authorized service center within the warranty period. 
 
To obtain warranty service, contact your nearest Fluke authorized service center 
or send the product, with a description of the difficulty, postage and insurance 
prepaid (FOB Destination), to the nearest Fluke authorized service center. Fluke 
assumes no risk for damage in transit. Following warranty repair, the product will 
be returned to Buyer, transportation prepaid (FOB Destination). If Fluke 
determines that the failure was caused by misuse, alteration, accident or 
abnormal condition of operation or handling,  Fluke will provide an estimate of 
repair costs and obtain authorization before commencing the work. Following 
repair, the product will be returned to the Buyer transportation prepaid and the 
Buyer will be billed for the repair and return transportation charges (FOB 
Shipping Point). 
 
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN 
LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING 
BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY 
OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE 
LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL 
DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING 
FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, 
RELIANCE OR ANY OTHER THEORY. 
 
Since some countries or states do not allow limitation of the term of an implied 
warranty, or exclusion or limitation of incidental or consequential damages, the 
limitations and exclusions of this warranty may not apply to every buyer. If any 
provision of this Warranty is held invalid or unenforceable by a court of 
competent jurisdiction, such holding will not affect the validity or enforceability of 
any other provision. 
 

 

Fluke Corporation  

Fluke Industrial B.V. 

 

P.O. Box 9090 

P.O. Box 90 

 

Everett WA 

7600 AB Almelo 

 

98206-9090, USA 

The Netherlands 

 

SERVICE CENTERS

 

To locate an authorized service center,  

visit us on the World Wide Web:  

http://www.fluke.com

  

or call Fluke using any of the phone numbers listed below: 

+1-888-993-5853 in U.S.A. and Canada 

+31-40-267-5200 in Europe 

+1-425-446-5500 from other countries 

 

®

i430flex

3000 A Flexible AC Current Probe 

Instruction Sheet 

Introducing the i430flex 

The i430flex is a flexible AC current probe for use with the Fluke  
430 series Power Quality Analyzers. The i430flex is optimized for 
current measurement on thick and hard to reach conductors. 
 

Unpacking 

The following items should be included in the package: 

  i430flex Current Probe(s)  

  Instruction Sheet (this paper) 

Check the contents of the shipping box for completeness. If 
something in the box has been damaged or missing, contact your 
distributor or the nearest FLUKE sales or service office immediately.
 

Safety Information 

 

 

Read First: Safety Information 

To ensure safe operation and service of the current 
clamp, follow these instructions: 

 

  Read the operating instructions before use and follow all safety 

instructions. 

  Use the Current Probe only as specified in the operating 

instructions; otherwise the probe’s safety features may not 
protect you. 

  Adhere to local and national safety codes. Individual protective 

equipment must be used to prevent shock and arc blast injury 
where hazardous live conductors are exposed. 

  Before each use, inspect the Current Probe and its latching 

system for any damage. Pay particular attention to the 
insulation surrounding the flexible measuring head. Look for 
cracks or missing portions of the probe housing or output cable 
insulation. Also look for loose or weakened components. 

  Do not use a probe that is cracked, damaged, or has a 

defective cable.  

  Never use the probe on a circuit with voltages higher than 

1000 V CAT III or 600V CAT IV 

 

CAT III equipment is designed to protect against transients 
in equipment in fixed equipment installations, such as 
distribution panels, feeders and short branch circuits, and 
lighting systems in large buildings. 

 

CAT IV equipment is designed to protect against transients 
from the primary supply level, such as an electricity meter 
or an overhead or underground utility service. 

 

 