i2000flex
2000A Flexible AC Current Probe
Instruction Sheet
Introducing the i2000flex
The i2000flex is a flexible AC current probe for use with
oscilloscopes and multimeters. The supplied Dual Banana to BNC
Adapter can be used for connection to instruments with banana
inputs. The i2000flex is optimized for current measurement on thick
and hard to reach conductors.
Unpacking
The following items should be included in the package:
•
i2000flex Current Probe
•
Dual Banana to BNC Adapter (model PM9081/001)
•
Instruction Sheet (this paper)
Check the contents of the shipping box for completeness. If
something in the box has been damaged or missing, contact your
distributor or the nearest FLUKE sales or service office
immediately.
Using the i2000flex Safely
Warning
To prevent electric shock, fire, and personal injury,
carefully read all safety information before
attempting to operate the Current Probe and follow
these procedures.
•
Before each use, inspect the Current Probe and its latching
system for any damage. Pay particular attention to the
insulation surrounding the flexible measuring head.
•
Do not use the Current Probe if it is damaged.
•
Do not use the Current Probe on circuits rated higher than
600V in Installation Category III (See "Specifications",
subsection “SAFETY”.)
•
Do not use the Current Probe to measure bare conductors
carrying a voltage from 30V up to 600V unless you are
wearing protective clothing suitable for high-voltage work.
Always use appropriate equipment for personal protection.
•
Do not expose the i2000flex unit to water.
In this Instruction Sheet, a Warning identifies conditions and
actions that pose hazard(s) to the user. A Caution identifies
conditions and actions that may damage the current probe.
International electrical symbols used are explained below.
See explanation in
the Instruction
Sheet
Conformité
Européenne
Double Insulation
(Protection class)
Specifications
SAFETY
i2000flex
(INPUT and OUTPUT)
Meets requirements of:
EN 61010-1,
UL3111,
for 600V
CAT III, pollution degree 2.
EMC
CE Conformity
Meets requirements of:
EN 50081-1
EN 50082-2
ELECTRICAL SPECIFICATIONS
All Electrical Specifications are valid at the following reference
conditions:
•
Ambient temperature
23±3°C (73±5.4°F)
•
Relative Humidity
0 to 85%
•
Frequency
48 to 65 Hz
•
Continuous external field
< 40 A/m
•
Load impedance on BNC
output
> 1 k
Ω
// < 100 pF
•
The current may not contain any DC component
•
No influence from adjacent currents
•
Conductor centered within the probe aperture
•
“POWER ON” indicator flashing (green LED)
“BATTERY LOW” indicator not flashing (red LED)
(see Figure 1)
Current Ranges
200A
2000A
Measuring range
2 to 200A ac
20 to 2000 A ac
Crest factor *)
up to 1.6
up to 1.6
Maximum non-
destructive current:
100 kA
100 kA
Output signal
10 mV/A
1 mV/A
Basic accuracy
±1% of full scale
±1% of full scale
Noise
< 2 mV rms
(200 mA rms)
< 2 mV rms
(2A rms)
Additional errors:
•
with temperature
(0 to +70 °C,
32 to 158 °F)
0.02% of
reading /°C,
0.036% of
reading /°F
0.02% of
reading /°C,
0.036% of
reading /°F
•
with position of
conductor in the
probe window
±2% of full scale
(bus
≥
2.5 cm or 1
inch from coupling)
±2% of full scale
(bus
≥
2,5 cm or 1
inch from coupling)
•
with external fields
(< 40 A/m)
±1% of full scale
(adjacent conductor
≥
20 cm or 8 inch
from head.)
±1% of full scale
(adjacent conductor
≥
20 cm or 8 inch
from head.)
Phase shift
50 to 60 Hz
See Figure 2
< ±0.5°
See Figure 3
< ±0.5°
*)
This is the maximum permissible ratio between the peak
value of a superimposed transient and the ac rms value.
Bandwidth (-3dB)
1 Hz to 20 kHz
Battery
9V alkaline IEC 6LR61,
550 mAhr
Battery life
> 100 hr
Indicators (see Figure 1)
Power On
Battery Low
Green LED flashing
Red LED flashing
GENERAL
Coupling diameter
∅
22.2 mm (0.874 in)
Weight
420 g (14.8 oz)
Transducer length
60 cm (31.5 in)
Transducer diameter
14.3 mm (0.563 in)
Minimum bending radius
38 mm (1.5 in)
Cable length from transducer to
box
1.8 m (71 in)
Cable length from box to BNC
output
0.5 m (19.7 in)
Maximum extension cable length
15 m (50 ft)
Weight (without battery)
420 gr (14.8 oz)
Temperature operating:
Transducer
i2000flex unit
-20 to +90°C (-4 to +194°F)
0 to + 70°C (32 to +158°F)
Temperature non-operating:
Transducer and
i2000flex unit
-40 to + 105°C (-40 to +221°F)
Relative Humidity
Operating
Non-operating
0 to 85%, < +35°C (+95°F)
0 to 75%, < +90°C (+194°F)
0 to 85%, < -10°C (+14°F)
0 to 75%, < +105°C (+221°F)
Altitude
Operating
Non-operating
to 3000 m (10000 ft)
to 12000 m (40000 ft)
Figure 1. Text on the i2000flex unit
PHASE SHIFT
200A Range
-14.0
-12.0
-10.0
-8.0
-6.0
-4.0
-2.0
0.0
2.0
1
10
100
1000
10000
100000
Frequency (Hz)
P
hase (D
egrees)
Figure 2. Phase Shift versus Frequency (200A range)
2000A Range
-14.0
-12.0
-10.0
-8.0
-6.0
-4.0
-2.0
0.0
2.0
1
10
100
1000
10000
100000
Frequency (Hz)
P
hase (D
egrees)
Figure 3. Phase Shift versus Frequency (2000A range)
*) Note:
When the battery is almost completely empty, none of
the LEDs will flash.
February 1998
© 1998 Fluke Corporation.
All rights reserved.
Printed in the Netherlands
All product names are trademarks of
their respective companies.
Current Ranges
{
Power off
Flashing when a
current range is
selected. *)
Flashing when the
battery is too
weak for correct
measurement
results. *)