Fluke i2000flex Instruction Sheet Download Page 1

®

i2000flex

2000A Flexible AC Current Probe

Instruction Sheet

Introducing the i2000flex

The i2000flex is a flexible AC current probe for use with
oscilloscopes and multimeters. The supplied Dual Banana to BNC
Adapter can be used for connection to instruments with banana
inputs. The i2000flex is optimized for current measurement on thick
and hard to reach conductors.

Unpacking

The following items should be included in the package:

• 

i2000flex Current Probe

• 

Dual Banana to BNC Adapter (model PM9081/001)

• 

Instruction Sheet (this paper)

 Check the contents of the shipping box for completeness. If
something in the box has been damaged or missing, contact your
distributor or the nearest FLUKE sales or service office
immediately.
 

 Using the i2000flex Safely

 

 Warning

 To prevent electric shock, fire, and personal injury,
carefully read all safety information before
attempting to operate the Current Probe and follow
these procedures.

 

• 

Before each use, inspect the Current Probe and its latching
system for any damage. Pay particular attention to the
insulation surrounding the flexible measuring head.

• 

Do not use the Current Probe if it is damaged.

• 

Do not use the Current Probe on circuits rated higher than
600V in Installation Category III (See "Specifications",
subsection “SAFETY”.)

• 

Do not use the Current Probe to measure bare conductors
carrying a voltage from 30V up to 600V unless you are
wearing protective clothing suitable for high-voltage work.
Always use appropriate equipment for personal protection.

• 

Do not expose the i2000flex unit to water.

 
 In this Instruction Sheet, a Warning identifies conditions and
actions that pose hazard(s) to the user. A Caution identifies
conditions and actions that may damage the current probe.
International electrical symbols used are explained below.
 

 

 See explanation in
the Instruction
Sheet

 

 Conformité
 Européenne

 

 Double Insulation
(Protection class)

 

 

 
 
 
 

 Specifications

 SAFETY

 

i2000flex
 (INPUT and OUTPUT)

 Meets requirements of:

 EN 61010-1,
UL3111,
for 600V
CAT III, pollution degree 2.

 EMC

 

 CE Conformity

 Meets requirements of:

 EN 50081-1
 EN 50082-2

 

 

 ELECTRICAL SPECIFICATIONS

 

All Electrical Specifications are valid at the following reference
conditions:

• 

Ambient temperature

 23±3°C (73±5.4°F)

• 

Relative Humidity

 0 to 85%

• 

Frequency

 48 to 65 Hz

• 

Continuous external field

 < 40 A/m

• 

Load impedance on BNC
output

 > 1 k

 // < 100 pF

• 

The current may not contain any DC component

• 

No influence from adjacent currents

• 

Conductor centered within the probe aperture

• 

“POWER ON” indicator flashing (green LED)
“BATTERY LOW” indicator not flashing (red LED)
(see Figure 1)

 

 Current Ranges

 

 200A

 2000A

 Measuring range

 2 to 200A ac

 20 to 2000 A ac

 Crest factor *)

 up to 1.6

 up to 1.6

 Maximum non-
destructive current:

 100 kA

 100 kA

 Output signal

 10 mV/A

 1 mV/A

 Basic accuracy

 ±1% of full scale

 ±1% of full scale

 Noise

 < 2 mV rms
(200 mA rms)

 < 2 mV rms
(2A rms)

 Additional errors:

 

 

• 

with temperature
(0 to +70 °C,
32 to 158 °F)

 0.02% of
  reading /°C,
 0.036% of
  reading /°F

 0.02% of
  reading /°C,
0.036% of
  reading /°F

• 

with position of
conductor in the
probe window

 ±2% of full scale
 (bus 

  2.5 cm or 1

inch from coupling)

 ±2% of full scale
 (bus 

 2,5 cm or 1

inch from coupling)

• 

with external fields
(< 40 A/m)

 ±1% of full scale
 (adjacent conductor

 20 cm or 8 inch

from head.)

 ±1% of full scale
 (adjacent conductor

 20 cm or 8 inch

from head.)

 Phase shift

 50 to 60 Hz

 See Figure 2
 < ±0.5°

 See Figure 3
 < ±0.5°

 

*)

This is the maximum permissible ratio between the peak

value of a superimposed transient and the ac rms value.

 

 

 Bandwidth (-3dB)

 

 1 Hz to 20 kHz

 Battery

 9V alkaline IEC 6LR61,
550 mAhr

 Battery life

 > 100 hr

 Indicators (see Figure 1)

 Power On
 Battery Low

 
 Green LED flashing
 Red LED flashing

 

 

 GENERAL

 

 Coupling diameter

 

 22.2 mm (0.874 in)

 Weight

 420 g  (14.8 oz)

 Transducer length

 60 cm (31.5 in)

 Transducer diameter

 14.3 mm (0.563 in)

 Minimum bending radius

 38 mm (1.5 in)

 Cable length from transducer to
box

 1.8 m (71 in)

 Cable length from box to BNC
output

 0.5 m (19.7 in)

 Maximum extension cable length

 15 m (50 ft)

 Weight (without battery)

 420 gr (14.8 oz)

 

 Temperature operating:

 Transducer
 i2000flex unit

 
 -20 to +90°C (-4 to +194°F)
 0 to + 70°C (32 to +158°F)

 Temperature non-operating:

 Transducer and
i2000flex unit

 
 -40 to + 105°C (-40 to +221°F)

 Relative Humidity

 Operating
 
 Non-operating
 

 
 0 to 85%, < +35°C (+95°F)
 0 to 75%, < +90°C (+194°F)
 0 to 85%, < -10°C (+14°F)
 0 to 75%, < +105°C (+221°F)

 Altitude

 Operating
 Non-operating

 
 to 3000 m (10000 ft)
 to 12000 m (40000 ft)

 

 

 

 Figure 1. Text on the i2000flex unit

 

 

 

 PHASE SHIFT

 

200A Range

-14.0

-12.0

-10.0

-8.0

-6.0

-4.0

-2.0

0.0

2.0

1

10

100

1000

10000

100000

Frequency (Hz)

P

hase (D

egrees)

 Figure 2. Phase Shift versus Frequency (200A range)

 

 

 

2000A Range

-14.0

-12.0

-10.0

-8.0

-6.0

-4.0

-2.0

0.0

2.0

1

10

100

1000

10000

100000

Frequency (Hz)

P

hase (D

egrees)

 Figure 3. Phase Shift versus Frequency (2000A range)

 

*) Note:

When the battery is almost completely empty, none of
the LEDs will flash.

 

 February 1998

 © 1998 Fluke Corporation.
 All rights reserved.
 Printed in the Netherlands
 All product names are trademarks of
their respective companies.