10.3.1.3
Resistance Inputs
Your SPRT or thermistor probe connects to a channel on the scanner using the
four-terminal binding posts. The top red and bottom black terminals (“C+” and
“C-”) source current to the probe. The middle two terminals (“P+” and “P-”)
sense the voltage on the probe. Figure 16 below shows how to connect a probe
115
10 2575 Scanner (optional)
Quick Start
Figure 15
1575 to 2575 Control Cable Connection