10.3
Quick Start
This section explains the basics of setting up and operating the 2575 Scanner
with the 1575 Thermometer. For more details on setting up the 1575, see Sec-
tion 4 starting on page 17.
10.3.1
Connections
The first step in operating the scanner with the 1575 Thermometer is to make
sure all the proper connections are made. Before connecting the cables between
the scanner and the 1575 Thermometer, switch the power off to the 1575.
10.3.1.1
Resistance Output
The output of the scanner must be properly connected to the input of the 1575.
The arrangement of the scanner Output terminals corresponds to that of the
1575 Probe 1 input terminals and should be connected accordingly as shown in
Figure 14 below.
10.3.1.2
Control Cable
Before connecting the control cable between the scanner and the 1575 Ther-
mometer, switch the power off to the 1575. The control input of the scanner
must be connected to the control output of the 1575 (see Figure 15). A 15-con-
ductor cable is provided for this. To ensure that the cable stays tightly con-
nected, fasten down the screws on the connectors.
1575/1590 Thermometer Readout
User’s Guide
114
Figure 14
Scanner Output Connection