The SECURITY soft-key allows the user to select which parameters and func-
tions are locked out. The security window appears as follows.
The INT REFR lock-out controls access to the SET INT and the CAL REFR
functions in the SYSTEM CAL menu. This locks out the internal reference re-
sistor calibration values, the oven, and the reference calibration procedure. This
group is locked out (ON) by default.
The PROBE,EXT lock-out controls access to the EDIT PROBE function in the
PROBE MENU and the SET EXT function in the SYSTEM CAL menu. This
locks out the probe selection, probe characterization and measurement parame-
ters, probe calibration functions, and the external reference value. This parame-
ter group is not locked out by default (OFF).
The SAMPLING lock-out controls access to the SAMPLE TIMING and the
DIGITAL FILTER functions in the SAMPLING MENU. This locks out the
sample timing parameters and the digital filter parameters. This parameter
group is not locked out by default (OFF).
The UNITS,TIME lock-out controls access to the UNITS and TIME functions
in the PARAM MENU. This locks out the units, time, and date settings. This
parameter group is not locked out by default (OFF).
The lock-out screen also allows the user to change the user password. The user
can enter a new password by typing a four-digit number in its place and press-
ing Enter. The number must be exactly four digits using the digits 0 through 9.
The decimal point, sign, and other keys cannot be used in the password. If the
password is changed, a message will appear and the user will be asked to con-
firm the operation.
1575/1590 Thermometer Readout
User’s Guide
96