Fluke-700LTP
Low-Pressure Test Pump
Instruction Sheet
Introduction
The Fluke-700LTP Low-Pressure Test Pump is a
portable, dual source of vacuum and pressure. Each
pump incorporates a vacuum and pressure selector
and fine adjustment control. The pump has the
following specifications:
•
Output pressure: 0 to 45 psi / 0 to 3 bar
•
Output vacuum: 0 to -13 psi / 0 to -900 mbar
•
Materials: Bright nickel-plated brass, clear
anodized aluminum
•
Adjustment: Fine volumetric pressure and
vacuum adjuster
•
Pressure Relief Valve: Adjustable from 0.7 to 45
psi / 50 mbar to 3 bar
•
Construction: Nylon, stainless steel, brass, PTFE,
and nitrile
•
Dimensions: 150 mm x 46 mm diameter
•
Weight: ~600 grams or 1.3 lbs (pump only)
Box Contents
•
Fluke-700LTP Low-pressure Test Pump
•
Two flexible hoses with two 1/4 in BSP female
and two 1/4 NPT female quick-fit connectors
•
Seal set
•
Adapter: one 1/4 in. NPT male to 1/8 in. NPT
male.
•
Instruction Sheet
How to Contact Fluke
To contact Fluke, call one of the following numbers:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Visit us on the World Wide Web at:
www.fluke.com
1
2
4
5
4
6
7
8
3
700ltp1f.eps
A
Release valve.
Use to reduce or release pressure in the
system. The rate of pressure reduction depends on the
degree of rotation when opening the valve. Minimal force
is required to seal the system.
B
Fine adjustment control.
Turn in or out to increase or
decrease pressure accordingly.
Caution
To avoid damage to the pump, do not wind the fine
adjustment any further when the top of the pump
body is visible.
C
Pressure/vacuum selection.
Press the selector as
indicated on the label to engage the desired mode. Make
sure that the release valve
A
is fully closed (clockwise
motion) prior to pumping.
D
Quick-fit connectors.
Fit the hose
E
and adapters
F
by screwing them into the connectors
D
by turning the
knurled nut on the connector fully counterclockwise.
E
Hose.
F
Quick-fit Adapters.
1/4 BSP female & 1/4 NPT female.
Use flat seals with BSP threads. Use PTFE with NPT
threads.
G
Pump Handle.
If the pump has not been used for a
while, there may be some resistance on the first stroke.
H
Pressure Relief Valve.
The maximum output pressure
can be set by adjustment of the pressure relief value
located inside the main piston. Access to the adjustment
screw is through the hollow-handle retaining screw.
Using a small screwdriver, turn adjustment screw
clockwise to increase pressure, and counterclockwise to
reduce pressure.