Fluke Fluke-700LTP Instruction Sheet Download Page 1

®

PN 1567810 June 2001, Rev.1, 8/02  

© 2001, 2002 Fluke Corporation. All rights reserved. Printed in U.S.A.

1

Fluke-700LTP

Low-Pressure Test Pump

Instruction Sheet

Introduction

The Fluke-700LTP Low-Pressure Test Pump is a
portable, dual source of vacuum and pressure. Each
pump incorporates a vacuum and pressure selector
and fine adjustment control. The pump has the
following specifications:

 

Output pressure: 0 to 45 psi / 0 to 3 bar

 

Output vacuum: 0 to -13 psi / 0 to -900 mbar

 

Materials: Bright nickel-plated brass, clear
anodized aluminum

 

Adjustment: Fine volumetric pressure and
vacuum adjuster

 

Pressure Relief Valve: Adjustable from 0.7 to 45
psi / 50 mbar to 3 bar

 

Construction: Nylon, stainless steel, brass, PTFE,
and nitrile

 

Dimensions: 150 mm x 46 mm diameter

 

Weight: ~600 grams or 1.3 lbs (pump only)

Box Contents

 

Fluke-700LTP Low-pressure Test Pump

 

Two flexible hoses with two 1/4 in BSP female
and two 1/4 NPT female quick-fit connectors

 

Seal set

 

Adapter: one 1/4 in. NPT male to 1/8 in. NPT
male.

 

Instruction Sheet

How to Contact Fluke

To contact Fluke, call one of the following numbers:

USA:  1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe:  +31 402-675-200
Japan:  +81-3-3434-0181
Singapore:  +65-738-5655
Anywhere in the world:  +1-425-446-5500

Visit us on the World Wide Web at:

www.fluke.com

1

2

4

5

4

6

7

8

3

700ltp1f.eps

A

Release valve. 

Use to reduce or release pressure in the

system. The rate of pressure reduction depends on the
degree of rotation when opening the valve. Minimal force
is required to seal the system.

B

Fine adjustment control. 

Turn in or out to increase or

decrease pressure accordingly.

Caution

To avoid damage to the pump, do not wind the fine
adjustment any further when the top of the pump
body is visible.

C

Pressure/vacuum selection. 

Press the selector as

indicated on the label to engage the desired mode. Make
sure that the release valve 

A

 is fully closed (clockwise

motion) prior to pumping.

D

Quick-fit connectors.  

Fit the hose 

E

 and adapters 

F

by screwing them into the connectors 

D

 by turning the

knurled nut on the connector fully counterclockwise.

E

Hose.

F

Quick-fit Adapters.

 1/4 BSP female & 1/4 NPT female.

Use flat seals with BSP threads. Use PTFE with NPT
threads.

G

Pump Handle.

 If the pump has not been used for a

while, there may be some resistance on the first stroke.

H

Pressure Relief Valve. 

The maximum output pressure

can be set by adjustment of the pressure relief value
located inside the main piston. Access to the adjustment
screw is through the hollow-handle retaining screw.
Using a small screwdriver, turn adjustment screw
clockwise to increase pressure, and counterclockwise to
reduce pressure.

Downloaded from

Arrow.com.