9
The Connector Adapter
ft/m/kft/km
Select feet, meters, kilofeet, or kilometers as the unit for length
measurements. The default is feet.
LOSS LIMIT
Sets the minimum level at which the troubleshooter shows the
warning
LOSS LIMIT
W
for an incident. For example, if you select
a limit of 2.0 dB, the troubleshooter shows the warning for
incidents above 2.0 dB. You can select a value from 0.5 dB (lower
loss) to 6.1 dB (higher loss) in 0.1 dB increments. The default is
0.5 dB.
REFLECTION LIMIT
Sets the minimum size at which the troubleshooter shows the
warning
REFLECTION LIMIT
W
for an incident. For example, if
you select a limit of -35 dB, the troubleshooter shows the
warning for incidents above -35 dB. You can select a value from
-24 dB (larger reflection) to -62 dB (smaller reflection) in 1 dB
increments. The default is -45 dB.
The Connector Adapter
You can change the SC connector adapter to connect to LC, ST,
or FC fiber connectors. See Figure 5.
W
Caution
Turn only the collar on the adapter. Do not use tools to
remove or install the adapters.