Innovative High Temperature Fiber Optic Infrared Pyrometers
Installation
31
one side and a RJ45 connector type on the counter side is used. Via the 4-conductor cable the
Endurance
®
-Series device can also be powered as a PD (Powered Device) by a PSE (Power
Sourcing Equipment) in a PoE (Power over Ethernet) mode. Refer to PoE standard IEEE
802.3af, mode A, 10/100 Mbit mixed DC & data.
Figure 19: M12 4-Conductor shielded cable with RJ45 on counter side
5.5.3.3. Endurance
®
specific terminal block
An Endurance
®
specific terminal block is available to attach the 12-wire color-coded sensor
cable via the terminal block to the process world.
Figure 20: Endurance
®
series labeled terminal block
5.5.4. Power Supply
Connections from a nominal 24VDC (500 mA or higher) power supply attach to the appropriate
terminals on the electronic enclosure’s terminal strip.
Isolated power is required, and the appropriate manufacturer supplied
power supply accessory provides this. Beware of use of other power
supplies, which may not provide the necessary isolation and could cause
instrument malfunction or damage!