![Fluke E1M User Manual Download Page 83](http://html1.mh-extra.com/html/fluke/e1m/e1m_user-manual_2309862083.webp)
Accessories
Electrical Accessories
83
12
12.1.1 High Temp 12-Wire Cable (E-2CCBxx)
Use the High Temp 12-wire cable for the sensor to support power supply, all inputs, outputs, and the RS485
interface. The cable described below is a shielded 12-conductor cable, made of two twisted pairs plus 8 separate
wires, all as tinned copper braid. The cable is equipped with a M16 DIN connector on one side and wire sleeves
at the counter side. The outer diameter of the cable is 7 mm (0.28 in).
The cable withstands ambient temperatures up to 200°C (392°F) and is Teflon coated. Teflon coated temperature
cables have good to excellent resistance to oxidation, heat, weather, sun, ozone, flame, water, acid, alkalis, and
alcohol, but poor resistance to gasoline, kerosene, and degreaser solvents.
Figure 12-1: 12-Wire Cable, High Temp Version
Table 12-1: Cable Specification
P/N
Ambient Temperature
Length
E-2CCB4
-80 to 200°C (-112°F to 392°F)
4 m (13 ft)
E-2CCB8
-80 to 200°C (-112°F to 392°F)
8 m (26 ft)
E-2CCB15
-80 to 200°C (-112°F to 392°F)
15 m (49 ft)
E-2CCB30
-80 to 200°C (-112°F to 392°F)
30 m (98 ft)
E-2CCB60
-80 to 200°C (-112°F to 392°F)
60 m (197 ft)
Color
Number
Cross Section
Shield
black/red
2 wires
0.33 mm² (AWG 22)
none
black/white
1 twisted pair
0.22 mm² (AWG 24)
yes
purple/gray
1 twisted pair
0.22 mm² (AWG 24)
yes
green/brown/blue/orange/yellow/clear
6 wires
0.22 mm² (AWG 24)
none
Table 12-2: Color Assignment to DIN Connector
Color
Connector Pin
Description
Black
A
RS485-A
White
B
RS485-B
Gray
C
- mA In
Purple
D
+ mA In
White/drain
E
Shield
Yellow
F
Trigger
Orange
G
Relay (alarm)
Blue
H
Relay (alarm)
Green
J
+ mA Out
Brown
K
- mA Out
Black
L
Ground (power)
Red
M
+ 24 VDC