![Fluke DTX-GFM2 User Manual Download Page 25](http://html1.mh-extra.com/html/fluke/dtx-gfm2/dtx-gfm2_user-manual_2309138025.webp)
Fiber Test Settings
19
Table 2. Fiber Test Settings (cont.)
Setting Description
SETUP > Fiber >
Index of Ref. Source (n)
> User Defined
or
Default
The tester uses the index of refraction (n) defined in the currently selected fiber type (
Default
) or
a value you define (
User Defined
). The default value defined by the selected fiber type
represents the typical value for that fiber type. You may enter a different value if necessary. To
determine the actual value, change the index of refraction until the measured length matches
the known length of a fiber.
Increasing the index of refraction decreases the reported length.
SPECIAL FUNCTIONS >
Set Reference
Patch Lengths
(softkey on the
View
Connections
screen)
Setting a reference sets the baseline power level for loss measurements. See “About Setting the
Reference” on page 11.
After you set the reference, you can enter the lengths of the reference test cords used. The
lengths are saved with results to meet TSB-140 reporting requirements for fiber test results.
Settings for saving tests See the
DTX Series CableAnalyzer Users Manual
or
Technical Reference Handbook
for details on
preparing to save tests.
Summary of Contents for DTX-GFM2
Page 15: ...Installing the Connector Adapter 9 amd99f eps Figure 4 Installing the Connector Adapter...
Page 31: ...Certifying Fiber Cabling 25 amd146f eps Figure 9 Smart Remote Mode Test Connections Method B...
Page 35: ...Certifying Fiber Cabling 29 amd147f eps Figure 11 Loopback Mode Reference Connections Method B...
Page 37: ...Certifying Fiber Cabling 31 amd148f eps Figure 12 Loopback Mode Test Connections Method B...
Page 45: ...Using the Visual Fault Locator 39 amd23f eps Figure 17 Using the Visual Fault Locator...
Page 60: ...DTX MFM2 GFM2 SFM2 Fiber Modules Users Manual B 4...