December 2017©2017 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies.
a3003 FC/a3004 FC
Wireless DC Clamp
Calibration Manual
Page 1: ...December 2017 2017 Fluke Corporation All rights reserved All product names are trademarks of their respective companies a3003 FC a3004 FC Wireless DC Clamp Calibration Manual...
Page 2: ...nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB D...
Page 3: ...ord 9 Performance Tests 10 Test the Display 10 Test the Backlight 10 Clamp Input and Keypad Test 10 DC Voltage Verification 11 4 20 mA Clamp Performance Test 12 2000 A Clamp Performance Test 13 2000 m...
Page 4: ...ii a3003 a3004 FC Calibration Manual...
Page 5: ...ating instructions see www fluke com Contact Fluke To contact Fluke use one of these telephone numbers Technical Support USA 1 800 44 FLUKE 1 800 443 5853 Calibration Repair USA 1 888 99 FLUKE 1 888 9...
Page 6: ...ts Have an approved technician repair the Product Use only specified replacement parts Comply with local and national safety codes Use personal protective equipment approved rubber gloves face protect...
Page 7: ...tation W WARNING RISK OF DANGER X WARNING HAZARDOUS VOLTAGE Risk of electric shock Do not apply around or remove from uninsulated hazardous live conductors without taking additional protective measure...
Page 8: ...a3003 FC a3004 FC Calibration Manual 4 Figure 1 a3003 FC Wireless DC Clamp Figure 2 a3004 FC Wireless DC Clamp 20 00A DC CU RR EN T CL AM P i20 00 Tactile Barrier Tactile Barrier...
Page 9: ...erature Operating Temperature 10 C 50 C Storage Temperature 40 C 60 C Temperature coefficient Add 0 1 X specified accuracy C 18 C or 28 C Operating Relative Humidity 90 at 35 C 75 at 40 C 45 at 50 C A...
Page 10: ...quipment is connected to a test object Korea KCC Class A Equipment Industrial Broadcasting Communication Equipment Class A Equipment meets requirements for industrial electromagnetic wave equipment an...
Page 11: ...to show The first digit flashes 3 Push and enter the first digit of the password Default is 1234 4 Push and enter the next digit of the password until all digits are entered 5 Push to confirm the inpu...
Page 12: ...t enclosure touching or handling any PCBs or components Minimize handling Handle static sensitive parts by non conductive edges Do not slide static sensitive components over any surface When removing...
Page 13: ...acement 2 Use a Phillips screwdriver to remove the bottom case screws Two of the screws are inside the battery door 3 Keep the PCA in the top case 4 Apply 3 0 V across the battery contacts on the PCA...
Page 14: ...2 Push while you keep pushed until all of the display segments are shown See Figure 4 If segments of the display are missing contact Fluke for repair See Contact Fluke Figure 4 Display Segments Test...
Page 15: ...y b Push The Meter is in verification mode 2 Connect the Calibrator to test leads and three of the six leads in the cable as shown in Figure 5 The white lead connects to the NORMAL Hi terminal of the...
Page 16: ...r the signal to stabilize d Record the reading Table 3 a3003 and a3004 DC Meter Performance Verification Test Connection Calibrator Output Resolution Specification UUT Reading Limits a3003 a3004 1 Yea...
Page 17: ...ode 4 Push on the Meter 5 For each value in Table 5 a Zero the Meter b Apply the Calibrator Output value to the Meter c Wait 4 seconds for the signal to stabilize d Record the reading All readings mus...
Page 18: ...ure 6 Calibration Connections for 2000 A Clamp 1 4 7 STBY 5522A CALIBRATOR HIGH CURRENT OUTPUT LO W CURRENT OUTPUT 850V PK MAX 850V PK MAX 30A PK 7V PK MAX 180A PK 7V PK MAX HI LO HI LO 3V PK 300mA PK...
Page 19: ...the performance tests do the calibration adjustment procedure Note The first calibration of the Product is available from the Fluke factory only Meter Calibration Adjustment Before you can adjust the...
Page 20: ...r 3 Use 1 loop insulated copper wire to short circuit the calibrator current output AUX and LO terminals see Figure 8 4 Clamp the 4 20 mA current clamp around the insulated copper wire Current flow sh...
Page 21: ...nction Step Display Calibrator Output mA DC 1 POS 20 mA 0 Hz 2 0A 0 mA 3 C 01 20 mA 0 Hz 4 0A 0 mA 5 C 02 20 mA 0 Hz 6 0A 0 mA 7 C 03 100 mA 0 Hz 8 0A 0 mA 9 C 04 100 mA 0 Hz 5522A 0 HI LO TC 1000V RM...
Page 22: ...til 0 A shows When 0 A shows on the display the balance adjustment passes Continue with the calibration To do the calibration adjustment 1 Move the jaw to the 3 kA coil 2 Power on all the equipment an...
Page 23: ...ssible damage to the Product or to equipment under test Remove batteries to prevent battery leakage and damage to the Product if it is not used for an extended period Be sure that the battery polarity...
Page 24: ...a3004 FC Calibration Manual 20 Replaceable Parts User replaceable parts and accessories are shown in Table 9 Table 9 User Replaceable Parts Description Fluke Part Number Battery Door 4583058 Batterie...