990DSL
Users Guide
4-24
If the fault is a short or if both conductors are faulted, or a shorting strap
has
already been placed at the far end
for a single-sided fault, the tester displays a
screen like Figure 4-13:
acy25s.eps
Figure 4-13. RFL Test Results: Double-sided Fault or Short
If your screen is similar to the one in Figure 4-13 and you
have not already
connected
a far-end strap, go to “Short or Double-Sided Fault Location Test
Results” for instructions on how to strap the pair and complete the test.
If your screen is similar to this and you
have already placed
a far-end T-R
shorting strap for location of a single-sided fault, press
D
(
Pre Strapped
Short (2w)
and continue below with step 2.
Note
The far end of a double-sided fault or short may be pre-strapped with
a good wire or pair prior to running the RFL test, with no change in
procedures.