Preparing the Instrument for Operation
Power Considerations
2
2-5
Power Considerations
The Instrument ships from the factory configured to match the requirements of your local
ac line power. If the Instrument is relocated to another region it may need to be
reconfigured to match the ac line power of the new location. Three things affect the
configuration:
•
Power cord (See Table 2-3.)
•
Line-power fuse (See Table 2-4 and Figure 2-1.)
•
Rear-panel switch setting (115-230, see Table 2-3 and Figure 2-1.)
The following paragraphs describe how to make the changes for a new voltage
configuration. They are also useful to verify that the Instrument’s current power
configuration is correct.
Replacing the Power Cord
XW
Warning
To avoid shock hazard, connect the instrument power cord to a
power receptacle that has an earth ground connection. A
protective ground connection by way of the grounding
conductor in the power cord is essential for safe operation.
The various power cords available for use with the Instrument are listed and shown in
Table 2-3. Use the table to identify your general location and the recommended LC
power cord. Match this LC number to one of the plugs in the drawing, and verify that the
plug on your power cable mates correctly with the local power outlets. If the plug is
incorrect, identify the correct LC number, and order the correct power cable from Fluke
using the part number from Table 2-3.
Table 2-3. Power Cord for Various Regions
Description Location Voltage/Current Part
Number
North America
120 V/15 A
LC1
284174
North America
240 V/15 A
LC2
2198736
Universal Euro
220 V/16 A
LC3
769422
United Kingdom
240 V/13A
LC4
769445
Switzerland
220 V/10 A
LC5
769448
China/Australia/New Zealand
240 V/10 A
LC6
658641
Power Cord
India/South Africa
240 V/5 A
LC7
782771
LC1
LC2
LC3
LC4
LC5
LC6
LC7
Ead05f.eps
Summary of Contents for 9640A Series
Page 8: ......
Page 16: ...9640A Instruction Manual viii...
Page 22: ...9640A Instruction Manual 1 2...
Page 40: ...9640A Instruction Manual 2 2...
Page 108: ...9640A Instruction Manual 4A 2...
Page 138: ...9640A Instruction Manual 4C 2...
Page 164: ...9640A Instruction Manual 4D 2...
Page 168: ...9640A Instruction Manual 4D 6...
Page 170: ...9640A Instruction Manual 4E 2...
Page 174: ...9640A Instruction Manual 4E 6...
Page 176: ...9640A Instruction Manual 4F 2...
Page 184: ...9640A Instruction Manual 4F 10...
Page 186: ...9640A Instruction Manual 4G 2...
Page 190: ...9640A Instruction Manual 4G 6...
Page 192: ...9640A Instruction Manual 5 2...
Page 272: ...9640A Instruction Manual 5 82...
Page 274: ...9640A Instruction Manual 6 2...
Page 284: ...9640A Instruction Manual 7 2...
Page 310: ...9640A Instruction Manual 8 2...
Page 316: ...9640A Instruction Manual 8 8...
Page 320: ...9640A Instruction Manual A 4...