HP 8662A/8663A Command Emulation
Emulated Commands
4F
4F-9
Request Service (RQS) Byte
Emulation mode attempts to recreate similar behavior of the SRQ and RQS feature of the
8662A/8663A. However, is not possible that the response timings will be the same, nor
will the timing for clearing bits within the SRQ.
The following register shows what bits are emulated by the Instrument
8 7 6
5 4
3
2
1
128 64
32 16
8 4 2 1
SP87 RQS Sweep
End
Param
Changed
Power
Fail
HW
error
Entry
Error
Ready
SP87
Sending this command causes an SRQ
Sweep End
When the last step of a sweep occurs
Param Change When any parameter of the output changes
Power Fail
When the generator is returned to a power ON condition (from standby
or off)
HW Error
When a hardware condition arises
Entry Error
when any invalid keystroke or program command occurs
Ready
When the generator is finished processing a Data message.
Summary of Contents for 9640A Series
Page 8: ......
Page 16: ...9640A Instruction Manual viii...
Page 22: ...9640A Instruction Manual 1 2...
Page 40: ...9640A Instruction Manual 2 2...
Page 108: ...9640A Instruction Manual 4A 2...
Page 138: ...9640A Instruction Manual 4C 2...
Page 164: ...9640A Instruction Manual 4D 2...
Page 168: ...9640A Instruction Manual 4D 6...
Page 170: ...9640A Instruction Manual 4E 2...
Page 174: ...9640A Instruction Manual 4E 6...
Page 176: ...9640A Instruction Manual 4F 2...
Page 184: ...9640A Instruction Manual 4F 10...
Page 186: ...9640A Instruction Manual 4G 2...
Page 190: ...9640A Instruction Manual 4G 6...
Page 192: ...9640A Instruction Manual 5 2...
Page 272: ...9640A Instruction Manual 5 82...
Page 274: ...9640A Instruction Manual 6 2...
Page 284: ...9640A Instruction Manual 7 2...
Page 310: ...9640A Instruction Manual 8 2...
Page 316: ...9640A Instruction Manual 8 8...
Page 320: ...9640A Instruction Manual A 4...