background image

9135

Infrared Thermometer Calibrator

User’s Guide

Rev. 5A2604

Hart Scientific

Summary of Contents for 9135

Page 1: ...9135 Infrared Thermometer Calibrator User s Guide Rev 5A2604 Hart Scientific...

Page 2: ...entific Division 799 E Utah Valley Drive American Fork UT 84003 9775 USA Phone 1 801 763 1600 Telefax 1 801 763 1010 E mail support hartscientific com www hartscientific com Subject to change without...

Page 3: ...cations 9 3 2 Environmental Conditions 9 3 3 Warranty 10 4 Quick Start 11 4 1 Unpacking 11 4 2 Set up 11 4 3 Power 11 5 Parts and Controls 13 5 1 Back Panel 13 5 2 Bottom 13 5 3 Top Panel 13 5 4 Right...

Page 4: ...9 Calibration Procedure 23 10 Maintenance 25 11 Troubleshooting 27 11 1 CE Comments 27 11 1 1 EMC Directive 27 11 1 2 Low Voltage Directive Safety 27 ii...

Page 5: ...he instrument or in this manual Symbol Description AC Alternating Current AC DC Battery CE Complies with European Union Directives DC Double Insulated Electric Shock Fuse PE Ground Hot Surface Burn Ha...

Page 6: ...to the user Caution identifies conditions and actions that may damage the instru ment being used 1 2 1 Warnings To avoid personal injury follow these guidelines DO NOT operate this unit without a pro...

Page 7: ...l to start a fire Operate the instrument in room temperatures between 5 40 C 41 104 F Allow sufficient air circulation by leaving at least three inches of space between the instrument and nearby objec...

Page 8: ...ail ServiceDesk fluke nl Fluke Int l Corporation Service Center Instrimpex Room 2301 Sciteck Tower 22 Jianguomenwai Dajie Chao Yang District Beijing 100004 PRC CHINA Phone 86 10 6 512 3436 Telefax 86...

Page 9: ...singa fluke com When contacting these Service Centers for support please have the following information available Model Number Serial Number Voltage Complete description of the problem 5 1 Before You...

Page 10: ...roportional integral tempera ture controller maintains the temperature of the plate The set point can be con veniently selected by the one button control to 50 C 100 C or 150 C One of three yellow LED...

Page 11: ...on Degree 2 per IEC1010 1 3 2 Environmental Conditions Although the instrument has been designed for optimum durability and trou ble free operation it must be handled with care The instrument should n...

Page 12: ...g To exercise this warranty the purchaser must forward the product after calling or writing a Hart Scientific Authorized Service Center see Section 1 3 for au thorization The Service Centers assume NO...

Page 13: ...ck the power connection 4 3 Power The calibrator is configured for either 115 VAC 10 50 60 Hz or 230 VAC 10 50 60 Hz using switches underneath There are two switches and both must be set to the correc...

Page 14: ...he mains supply must be reasonably stable and free from surges dropouts and interference that can affect the temperature stability of the IR target 9135 Infrared Thermometer Calibrator User s Guide 12...

Page 15: ...1 IR Target The large hole in the top panel gives access to the IR target The IR target is a metal plate that is coated with a special paint with an emmisivity of 0 95 The IR target is controlled at t...

Page 16: ...e the 9135 against a standard RTD see Section 9 Calibration Access Hole There is a small oblong hole on the right side panel These gives access to two potentiometers used to calibrate the 9135 DO NOT...

Page 17: ...entiometers unless qualified to do so as part of a regular mainte nance and calibration schedule see Section 9 15 5 Parts and Controls Right Side Panel Prop Calibration Access Hole Test Well Figure 2...

Page 18: ...erature stabilizes To select the 100 C set point press the select button once and observe the 100 C set point LED light up To select the 150 C set point press the select button once more and observe t...

Page 19: ...s the temperature of the IR target using a thermistor sensor imbedded in the plate It compares the resistance of the thermistor to preset set point resistances to determine whether the IR target shoul...

Page 20: ...d When calibrating IR thermometers the thermometer must see only the IR target of the 9135 The thermometer must be pointed directly at the tar get at approximately a 90 angle and with a close enough p...

Page 21: ...peration of the 9135 and especially affect the ability to calibrate the instrument in the future DO NOT attempt to insert fluid grease or any objects other than clean thermometers into the well Insert...

Page 22: ...shows the temperature is close to 50 C within 0 2 C Adjust the potentiometer clock wise to increase the temperature or counter clockwise to decrease the temperature 4 Set the set point to 150 C and l...

Page 23: ...care Avoid knocking bumping or dropping the calibrator Due to the size and portability of the 9135 it is vulnerable to being shocked and dropped as it is carried around It might be inadvertently drag...

Page 24: ...tor cause loss of accuracy and may cause sticking and damage to your probes In the event that the calibrator should require service or repair please contact a Hart Scientific Authorized Service Center...

Page 25: ...correct voltage as indicated on the back of the calibrator The internal fuse may be blown or the calibrator may be damaged in which case the instrument should be returned Contact a Hart Scien tific Au...

Reviews: