9118A
Users Manual
62
Prepare Test Equipment and Power
On
Set setpoint
Heat On
Take As Found Data
for stability
Pass/Fail
Take As Found Data for
axial uniformity
Calculate the difference
of +/- 30 mm
Set the unifomity align
parameter
Take As Found Data
for stability
Pass/Fail
Take As Found Data for
Axial Uniformity
Calculate the gradient
Set the gradient align
parameter
Calculate the accuracy
Set the accuracy
parameter
Take As Found Data
for stability
Take As Found Data
for Axial Gradient
Take As Found Data
for Accuracy
Finished
Calculate the uniformity
Test the stabiltiy
All
Calibration Point
Finished
Yes
No
Fail
Fail
Fail
Fail
Fail
Fail
Fail
Fail
Pass
Pass
Pass
Pass
Pass
Pass
Take As Found Data
for stability
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
gxs114.eps
Figure 24. Tube Furnace Calibration Flow Chart