under the unit and carefully lifting at the same time. Never move the fur-
nace if it is hot.
•
Air circulated through the gap surrounding the furnace core keeps the
chassis cool.
DO NOT SHUT OFF THE FURNACE WHILE AT
HIGH TEMPERATURES.
The fan will turn off allowing the chassis to
become hot. Alternatively, if used, the cooling water should remain on un-
til the furnace is cool.
•
Once the unit has been taken to high temperatures (over 800°C), it takes
days for the unit to cool completely.