![Fluke 9100A Series Manual Download Page 40](http://html.mh-extra.com/html/fluke/9100a-series/9100a-series_manual_2309313040.webp)
9100A-017
4-16
becomes critical as clock speeds increase above 5 Mhz. This
duty cycle ensures that the the minimum clock pulse width of 50
ns is maintained. If the clock does not have a duty cycle close to
50%, the external sync mode is recommended instead of the
capture mode for high speeds.
When capture clock is used, performing the external calibration
procedure (described earlier in this section) is recommended. If
a UUT has considerable propagation delay, the TL/1
setoffset
command can adjust for this. (The sync mode must be set to
“capture” prior to using the
setoffset
command.)
SYNCHRONIZING MULTIPLE MODULE VECTORS
4.17.
To synchronize the output vectors of multiple modules, the same
clock source must be applied to the DR CLK input of all the
modules that are to be synchronized. The INT CLK clock
(selectable at 1, 5, 10, or 20 MHz) may be used as the clock
source by connecting it to the DR CLK input of all modules. An
external clock at any frequency may also be used (external clock
signals must meet the specifications listed in Section 2).
As the speed of the clock increases, high-speed signal tech-
niques must be used to ensure signal integrity. A connector made
of high-quality cabling with each set of signal and ground lines
the same length gives good results. In some cases, providing a
separate buffered clock signal to each DR CLK input may be
necessary.
Summary of Contents for 9100A Series
Page 6: ...vi ...
Page 8: ...viii ...
Page 10: ...x ...
Page 14: ...9100A 017 1 4 ...
Page 24: ...9100A 017 3 6 ...
Page 44: ...9100A 017 5 4 ...
Page 58: ...9100A 017 6 14 ...
Page 83: ...A 1 Appendix A New TL 1 Commands ...
Page 84: ...9100A 017 A 2 ...
Page 87: ...clockfreq 3 For More Information The Overview Of TL 1 section of the Programmer s Manual ...
Page 88: ...clockfreq 4 ...
Page 91: ...drivepoll 3 For More Information The Overview Of TL 1 section of the Programmer s Manual ...
Page 92: ...drivepoll 4 ...
Page 104: ...vectordrive 4 ...
Page 107: ...vectorload 3 For More Information The Overview Of TL 1 section of the Programmer s Manual ...
Page 108: ...vectorload 4 ...
Page 116: ...9100A 017 C 2 ...
Page 117: ...9100A 017 C 3 ...
Page 118: ...9100A 017 C 4 ...