9100A-017
3-3
transition has occurred, vector driving resumes on the next
qualified clock. The edge polarity cannot be changed from one
polarity to the other within the vector file.
STOPPING VECTOR DRIVING
3.7.
Vector driving is stopped if the STOP statement is reached in a
vector file, if the programmed edge on the external STOP line is
received, or if a
writeword
or
writepin
command is performed
during driving. Vector driving is suspended if the WAIT state-
ment has been reached within the vector file, if the external
ENABLE condition set by
enableoutput
is not satisfied, or if no
clock for the appropriate
syncoutput
mode is received.
Figure 3-1. WAIT Input Timing Diagram