8845A/8846A
Users Manual
4-18
1.
Press
L
to display the measurement setup menu.
2.
Press the
TRIGGER
soft key.
3.
Press the soft key labeled
#SAMPLES
.
4.
Use the soft keys to set the number of samples between 1 and 50,000.
Select the digit in the value by pressing either
<--
or
-->
.
With the digit selected, press the soft key labeled
--
to decrement the digit or
++
to
increment the character.
5.
Press
ENTER
.
Connecting an External Trigger
The TRIG I/O jack on the Meter’s rear panel is used to connect to an external trigger
signal. The falling edge of a TTL signal will trigger the Meter to start taking
measurements, if the Meter is in external trigger mode.
Figure 4-9 indicates the purpose of the three pins in the TRIG I/O connector.
TRIG I/O
Pin 3: Measure Complete Out
Pin 1: Trigger In
Pin 2: Signal Ground
caw059.eps
Figure 4-9. TRIG I/O Pin-out Description
Monitoring the Measurement-Complete Signal
In addition to being a trigger input, the TRIG I/O jack on the Meter’s rear panel also
provides a signal indicating the completion of a measurement cycle. A falling edge of a
TTL signal indicates a measurement cycle is complete. See Figure 4-9 above to identify
which pins on the TRIG I/O connector are used to sense the measurement-complete
signal.
Summary of Contents for 8845A
Page 6: ...8845A 8846A Users Manual iv...
Page 8: ...8845A 8846A Users Manual vi...
Page 10: ...8845A 8846A Users Manual viii...
Page 12: ...8845A 8846A Users Manual 1 2 Diode Test 1 22 Measurement Rates IEEE488 4 1 22...
Page 34: ...8845A 8846A Users Manual 2 2...
Page 42: ...8845A 8846A Users Manual 2 10...
Page 70: ...8845A 8846A Users Manual 4 2...
Page 88: ...8845A 8846A Users Manual...
Page 90: ...8845A 8846A Users Manual A 2...
Page 96: ...8845A 8846A Users Manual C 2...
Page 102: ...8845A 8846A Users Manual 4...