Accuracy (8846A)
Accuracy is given as
±
(% measu % of range)
Range
24 Hour
(23
±
1
°
C)
90 Days
(23
±
5
°
C)
1 Year
(23
±
5
°
C)
Temperature
Coefficient/
°
C
Outside 18 to 28
°
C
100
μ
A
0.01 + 0.02
0.04 + 0.025
0.05 + 0.025
0.002 + 0.003
1 mA
0.007 + 0.005
0.030 + 0.005
0.05 + 0.005
0.002 + 0.0005
10 mA
0.007 + 0.02
0.03 + 0.02
0.05 + 0.02
0.002 + 0.002
100 mA
0.01 + 0.004
0.03 + 0.005
0.05 + 0.005
0.002 + 0.0005
400 mA
[3]
0.03 + 0.004
0.04 + 0.005
0.05 + 0.005
0.005 + 0.0005
1 A
[2]
0.03 + 0.02
0.04 + 0.02
0.05 + 0.02
0.005 + 0.001
3 A
[1][2]
0.05 + 0.02
0.08 + 0.02
0.1 + 0.02
0.005 + 0.002
10 A
[2]
0.1 + 0.008
0.12 + 0.008
0.15 + 0.008
0.005 + 0.0008
[1] Part of 10 A range
[2] Available at front panel connectors only
[3] 400 mA available in software version 2.0 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1
minute off.
Accuracy (8845A)
Accuracy is given as
±
(% measu % of range)
Range
24 Hour
(23
±
1
°
C)
90 Days
(23
±
5
°
C)
1 Year
(23
±
5
°
C)
Temperature
Coefficient/
°
C
Outside 18 to 28
°
C
100
μ
A
0.01 + 0.02
0.04 + 0.025
0.05 + 0.025
0.002 + 0.003
1 mA
0.007 + 0.005
0.030 + 0.005
0.05 + 0.005
0.002 + 0.0005
10 mA
0.007 + 0.02
0.03 + 0.02
0.05 + 0.02
0.002 + 0.002
100 mA
0.01 + 0.004
0.03 + 0.005
0.05 + 0.005
0.002 + 0.0005
400 mA
[3]
0.03 + 0.004
0.04 + 0.005
0.05 + 0.005
0.005 + 0.0005
1 A
[2]
0.03 + 0.02
0.04 + 0.02
0.05 + 0.02
0.005 + 0.001
3 A
[1][2]
0.05 + 0.02
0.08 + 0.02
0.10 + 0.02
0.005 + 0.002
10 A
[2]
0.10 + 0.008
0.12 + 0.008
0.15 + 0.008
0.005 + 0.0008
[1] Part of 10 A range
[2] Available at front panel connectors only
[3] 400 mA available in software version 2.0 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1
minute off.
Additional Current Errors
Digits NPLC
Additional NPLC Noise Error for
1 mA, 100 mA, 400 mA, 3 A and
10 A
Additional NPLC Noise Error
for 100
μ
A, 10 mA, 1 A
6½
100
0 % of range
0 % of range
6½
10
0 % of range
0 % of range
5½
1
0.001 % of range
0.01 % of range
5½ 0.2
0.011 % of range
±
4
μ
A
0.11 % of range
±
4
μ
A
4½ 0.02
0.04 % of range
±
4
μ
A
0.28 % of range
±
4
μ
A
AC Current
The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs
from 1 % to 5 % of range, add an additional error of 0.1 % of range.
Input Protection....................................................... Tool accessible 11 A/1000 V and 440 mA/1000 V fuses, limits of
400 mA continuous 550 mA for 2 minutes on, 1 minute off.
Measurement Method ............................................. ac-coupled true-rms, dc-coupled to the fuse and shunt (no blocking
capacitor)
AC Filter Bandwidth
Slow .................................................................... 3 Hz to 10 kHz
Medium ............................................................... 20 Hz to 10 kHz
Summary of Contents for 8845A
Page 6: ...8845A 8846A Users Manual iv...
Page 8: ...8845A 8846A Users Manual vi...
Page 10: ...8845A 8846A Users Manual viii...
Page 12: ...8845A 8846A Users Manual 1 2 Diode Test 1 22 Measurement Rates IEEE488 4 1 22...
Page 34: ...8845A 8846A Users Manual 2 2...
Page 42: ...8845A 8846A Users Manual 2 10...
Page 70: ...8845A 8846A Users Manual 4 2...
Page 88: ...8845A 8846A Users Manual...
Page 90: ...8845A 8846A Users Manual A 2...
Page 96: ...8845A 8846A Users Manual C 2...
Page 102: ...8845A 8846A Users Manual 4...