![Fluke 7526A Calibration Manual Download Page 21](http://html1.mh-extra.com/html/fluke/7526a/7526a_calibration-manual_2310418021.webp)
Precision Process Calibrator
Performance Tests
13
DC Voltage
The dc voltage performance test verifies the accuracy of the
Volts Source
output from
the front panel of the Product. Verify the measurements shown in Table 4 with a digital
multimeter (DMM).
Table 4. DC Voltage Verification
Range
Nominal Value
Measured Value
90-Day
Specification
(V)
1-Year
Specification
(V)
100 mV
0 V
3.00E-06
3.00E-06
0.035 V
3.88E-06
4.05E-06
0.065 V
4.63E-06
4.95E-06
0.1 V
5.50E-06
6.00E-06
1.0 V
0 V
1.00E-05
1.00E-05
0.35 V
1.88E-05
2.05E-05
0.65 V
2.63E-05
2.95E-05
1 V
3.50E-05
4.00E-05
10.0 V
0 V
1.00E-04
1.00E-04
3.5 V
1.88E-04
2.05E-04
6.5 V
2.63E-04
2.95E-04
10 V
3.50E-04
4.00E-04
100.0 V
0 V
1.00E-03
1.00E-03
35 V
1.88E-03
2.05E-03
65 V
2.63E-03
2.95E-03
100 V
3.50E-03
4.00E-03
Summary of Contents for 7526A