74X Series
Calibration Manual
28
Loop Power
Proceed as follows to verify the Loop Power function.
1.
Connect the UUT to the HP3458A DMM as shown in Figure 14.
2.
On the UUT press [SETUP], [ENTER], select
Loop Power 24V
, and press [ENTER]
again.
3.
Observe the no-load voltage reading on the DMM and verify that it is within the
range shown in Table 18.
4.
Set the Loop Power voltage to the 28V setting and repeat step 3.
5.
When complete, disable Loop Power through the [SETUP] menu or by turning the
UUT off. This saves battery life.
TC
TC
mA
mA
V
V
RTD
RTD
SOURCE
300V
MAX
30V
MAX
30V
MAX
30V
MAX
MEAS
CAT
FLUKE 74X
Ω
SENSE
INPUT
HI
LO
GUARD
HI
LO
HP3458A
gb15f.eps
Figure 14. Loop Power Verification Connections
Table 18. Loop Power Verification
UUT SETTING
Output Accuracy 1 and 2Year
24 Volt
22.8 to 25.2
28 Volt
26.6 to 29.4