Performance Verification Tests
27
RTD Source
Proceed as follows to verify the RTD Source function:
1.
Connect the UUT to the HP3458A DMM as shown in Figure 9. Use a 4-wire
connection transitioning to 2-wires at the UUT. Set the HP3458A to 4-Wire Ohms,
auto-range.
2.
Set the UUT to the RTD Source function, Pt100 (385) at -180
°
C, ITS-90 scale.
3.
Observe the resistance reading on the DMM and check to see if it is within the range
shown under the appropriate column in Table 17.
4.
Continue through the test points.
5.
When you are finished, press
c
twice on the UUT to set the Source function to off.
This saves battery life.
Table 17. RTD Source Verification (100W Pt (385))
UUT
Output
Nominal
(Ohms)
Minimum
1 Year
Maximum
1 Year
Minimum
2 Year
Maximum
2 Year
-180
°
C
27.096
27.053
27.139
27.011
27.181
100
°
C
138.505
138.429
138.581
138.353
138.657
780
°
C
369.712
369.592
369.832
369.562
369.862