Fluke 700HTP-2 Instruction Sheet Download Page 1

PN 4096896 
September 2011 

2011 Fluke Corporation. All rights reserved. Printed in U.S.A. 

All product names are trademarks of their respective companies. 

 

 

How to Prime the Test Pump 

Before you connect the Pump to the Unit Under Test (UUT), the 
Pump must be filled with fluid and primed for use. Follow the 
directions below to prepare the Pump for use. 

1.  Fill the Pump with fluid: 

a.  Remove the Reservoir Filler Cap (item 

.) 

b.  Fill the Hydraulic Fluid Reservoir approximately two-thirds full 

(66 cc) with distilled water, mineral oil, or hydraulic oil.  

Note 

If water is used in the Pump, use only DISTILLED water 
and drain the water out of the Pump after use.  

c.  Install the Reservoir Filler Cap (item 

.) Tighten by hand. 

2.  Prime the Pump: 

a.  Adjust the Vernier Fine Adjustment Knob (item 

) to the 

approximate middle-range.

 

b.  Turn the Pressure Bleed Knob (item 

) clockwise until it 

stops in the full closed position.  

c.  Squeeze the handles together to pump fluid. Pump fluid until 

fluid shows at the Pump Port. 

d.  Attach the fluid hose to the Pump. 

e.  Squeeze the handles together until fluid shows at the end of 

the hose. 

f.  Attach the fluid hose to the Unit Under Test (UUT) and 

squeeze the Pump handles until pressure increases.  

g.  Bleed the trapped air from the Pump. Pressurize the Pump to 

1,000 psi then slowly turn the Pressure Bleed Knob (item 

counter-clockwise to release pressure. 

Notes 

 

As the pressure rises it will be hard to squeeze the 
handles together. Use the Handle Limiter (item 

) to 

limit the travel of the handles which will make them 
easier to squeeze together. To toggle the Handle 
Limiter on or off, squeeze the handles together and 
push the Handle Limiter (item 

) in or out.  

 

The rate at which the pressure increases is directly 
related to the test configuration (hose length and 
volume of UUT.)  

Replacement Parts 

Rebuild Kit, Fluke PN 4097836

 

Hydraulic Fluid Reservoir Rebuild Kit, Fluke PN 2844341 
Handle Assembly Rebuild Kit, Fluke PN 2844352 

LIMITED WARRANTY AND LIMITATION OF LIABILITY

 

This Fluke product will be free from defects in material and workmanship for 
one year from the date of purchase. This warranty does not cover fuses, 
disposable batteries, or damage from accident, neglect, misuse, alteration, 
contamination, or abnormal conditions of operation or handling. Resellers are 
not authorized to extend any other warranty on Fluke’s behalf. To obtain 
service during the warranty period, contact your nearest Fluke authorized 
service center to obtain return authorization information, then send the product 
to that Service Center with a description of the problem. 
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, 
SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR 
IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, 
INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING 
FROM ANY CAUSE OR THEORY. Since some states or countries do not 
allow the exclusion or limitation of an implied warranty or of incidental or 
consequential damages, this limitation of liability may not apply to you. 

11/99 

 

 

Fluke-700HTP-2 

Hydraulic Test Pump 

Instruction Sheet

 

 

 

Introduction 

The Fluke 700HTP-2 Hydraulic Test Pump (the Pump or Product) is a 
portable source of high pressure.  

Safety 

W

 

Warning 

To prevent personal injury, use the Product only as 
specified, or the protection supplied by the Product can be 
compromised. 

Pump Specifications 

  Maximum pressure: 10,000 psi (690 bar) 

  Hydraulic media: distilled water, mineral based hydraulic oil, or up 

to hydraulic oil 100 (30 W). 

  Wetted materials: 303 stainless steel, aluminum, polyurethane, 

PTFE, nitrile, and nylon. 

  Weight: 1.29 lb (0.58 kg) 

  Dimensions: length 9.5 in (241 mm), width 6.1 in (155 mm), depth 

2.6 in (66 mm) 

Box Contents 

  Model 700HTP-2 Hydraulic Test Pump with a 1/4-inch NPT tee 

installed. 

  Two 1/4-inch NPT male to 1/4-inch BSP female adapters. 

  One 1/4-inch NPT male to 1/4-inch NPT male fitting. 

  One 1/4-inch NPT male to 1/8-inch NPT male fitting. 

  Instruction Sheet 