Fluke 5623B User Manual Download Page 12

5623B

Users Guide

6

Troubleshooting

In the event that the probe appears to function abnormally, this section may be of use in solving 

the problem. Several possible problem conditions are described along with likely causes and 

solutions. If a problem arises, please read this section carefully and attempt to understand and 

solve the problem. If the probe seems faulty or the problem cannot otherwise be solved, contact 

Fluke for assistance (see “How to Contact Fluke”). Be sure to have the model number and serial 

number of your probe available.

Table 3. PRT Troubleshooting 

Problem

Solution

Data changes greater than 0.1 °C 

are observed.

Mechanical shock can cause temperature errors as great as 0.5 °C. If 

this is observed, first measure and record the R

tpw

.

Data changes less than 0.1 °C.

Slight mechanical shock can cause temperature errors less than 0.1 °C.

Data unstable.

• 

If the data is unstable at the Triple Point of Water (TPW), check 

connections for evidence of a bad connection. If the connector 

appears to be in good condition and the connections are good, the 

PRT may be damaged. 

Contact Fluke for assistance (see “How to 

Contact Fluke”).

• 

If the data is unstable at high temperatures, it may be due to electrical 

noise in the system. Reduce the temperature and observe the data. 

If it is stable, electrical noise is interfering with the measurements at 

high temperatures. Check the grounding of the readout device and 

the heat source. A faulty ground on either device could interfere with 

high temperature measurements. A ground wire attached to the metal 

sheath of the PRT may help to reduce electrical noise interference.

Temperature readout is different 

than expected. For example, the 

heat source is set at 150 °C, the 

PRT measures 125 °C.

• 

Measure the PRT resistance at TPW.

• 

The PRT sensor may be shorted if the measured resistance at 0 °C is 

significantly less than the probe’s nominal resistance. For example, a 

probe whose nominal resistance is 100 

Ω

 but measures 70 

Ω

 at 0 °C. 

•  If the resistance of the PRT is only a few ohms, there may be a short 

in the four lead-wires.

•  If the PRT is open, the resistance will be “Out of Limits” or in the 

kilohm or megohm range.

For more information or assistance, 

Contact Fluke (see “How to 

Contact Fluke”).

Summary of Contents for 5623B

Page 1: ...1 Fluke Corporation All rights reserved Specifications subject to change without notice Users Guide All product names are trademarks of their respective companies Test Equipment Depot 800 517 8431 99...

Page 2: ......

Page 3: ...Recalibration 1 Before You Start 1 Symbols Used 1 Safety Information 2 How to Contact Fluke 2 Specifications 3 Installation 4 Lead Wire Identification 4 PRT Care and Handling Guidelines 4 Operation 5...

Page 4: ...5623B Users Guide ii...

Page 5: ...iii List of Tables Table Title Table 1 International Electrical Symbols 1 Table 2 Specifications 3 Table 3 PRT Troubleshooting 6 Page List of Figures Figure Title Page Figure 1 PRT Schematic 4...

Page 6: ...5623B Users Guide iv...

Page 7: ...ed For information on recalibrating your 5623B contact Fluke for assistance see How to Contact Fluke Depending on the user s Quality Assurance requirements the PRT drift should be checked periodically...

Page 8: ...periods of time can cause the handle to become hot Follow all safety guidelines listed in the Users Guide Calibration Equipment should only be used by trained personnel Use the Product only as specif...

Page 9: ...Sheath Dimensions Length x Diameter 152 mm x 6 35 mm 6 in x 0 25 in Sheath Material Inconel 600 Transition Junction Temperature Range 2 100 C to 156 C Minimum Immersion Length 3 5 mK error 114 mm 4 5...

Page 10: ...Guidelines Keep the thermometer as clean as possible Immerse the thermometer in the appropriate liquid for the temperature range If a dry block is used the well diameter should allow the PRT to comfor...

Page 11: ...th required can be determined by performing a gradient test taking measurements approximately every 1 27 cm 0 5 in until there is a significant difference in readings Allow the thermometer to stabiliz...

Page 12: ...ood condition and the connections are good the PRT may be damaged Contact Fluke for assistance see How to Contact Fluke If the data is unstable at high temperatures it may be due to electrical noise i...

Reviews: