Calibration and Verification
Performance Verification Tests
3
3-37
3-46. AC Voltage Amplitude Accuracy, Squarewave (AUX)
The AC Voltage Amplitude Accuracy, Squarewave (AUX) test checks the amplitude
accuracy at the AUX terminals. For this test, use the Fluke 5790A. Refer to the 5790A
Operator Manual for operating instructions and connections. For squarewaves, the
measured value (in rms) should be exactly 1/2 the nominal value in peak-to-peak. Table
3-33 shows the test points.
Table 3-33. AC Voltage Amplitude Accuracy, Squarewave (AUX)
Nominal Value
(p-p, NORMAL)
Nominal Value
(p-p, AUX)
Frequency
Measured Value
(V rms, AUX)
Deviation %
1-Year Spec.
(%)
3 V
300 mV
10 Hz
1.350
3 V
300 mV
1 kHz
0.800
3 V
300 mV
5 kHz
6.100
3 V
300 mV
10 kHz
6.100
3 V
3 V
10 Hz
1.350
3 V
3 V
1 kHz
0.800
3 V
3 V
5 kHz
6.100
3 V
3 V
10 kHz
6.100
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......