5500A
Service Manual
3-10
Table 3-8. Resistance Calibration Steps
Step
5500A Output
Comments
1
1
Ω
Make a four-wire measurement
2
10
Ω
“
3
11
Ω
“
4
32
Ω
“
5
35
Ω
“
6
100
Ω
“
7
110
Ω
“
8
320
Ω
“
9
0.35 k
Ω
“
10
1 k
Ω
“
11
1.1 k
Ω
“
12
3.2 k
Ω
“
13
3.5 k
Ω
“
14
10 k
Ω
“
15
11 k
Ω
“
16
32 k
Ω
“
17
35 k
Ω
“
18
100 k
Ω
“
19
110 k
Ω
Make a two-wire measurement
20
320 k
Ω
“
21
0.35 M
Ω
“
22
1 M
Ω
“
23
1.1 M
Ω
“
24
3.2 M
Ω
“
25
3.5 M
Ω
“
26
10 M
Ω
“
27
11 M
Ω
“
28
32 M
Ω
“
29
35 M
Ω
“
30
100 M
Ω
“
31
110 M
Ω
“
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......