5500A
Service Manual
6-112
6-137. Equipment Required
The following equipment is necessary for performing the hardware adjustments
described in this section. The models listed are recommended for providing accurate
results.
•
Standard adjustment tool for adjusting the pots and trimmer caps
•
Extender Card (pn 661865, 5800A-7006K, Extender Kit )
•
Oscilloscope Mainframe and Sampling Head (Tektronix 11801 with SD-22/26 or
Tektronix TDS 820 with 8 GHz bandwidth)
•
10 dB Attenuator (Weinschel 9-10 (SMA), or Weinschel 18W-10, or equivalent)
•
Cable provided with SC300
•
Spectrum Analyzer (Hewlett-Packard 8590A)
6-138. Adjusting the Leveled Sine Wave Function
There is one adjustment procedure that needs to be made for the leveled sine wave
function. The procedure adjusts the harmonics.
6-139. Equipment Setup
This procedure uses the spectrum analyzer. Before you begin this procedure, verify that
the Calibrator Mainframe is in leveled sine wave mode (the Levsine menu is displayed),
and program it to output 5.5 V p-p @ 50 MHz. Press
O
to activate the output.
Refer to Figure 6-24 for setup connections and connect the Calibrator Mainframe to the
Spectrum Analyzer. Adjust the Spectrum Analyzer so that it displays one peak across its
horizontal centerline. The far right of the peak is fixed at the far right of the centerline,
as shown below.
6-140. Adjusting the Leveled Sine Wave Harmonics
Note
This procedure should only be used for adjusting the leveled sine wave
harmonics. Do not use this procedure as a verification test. The
specifications in this procedure are not valid for verification.
Set the Spectrum Analyzer to the parameters listed below.
Spectrum Analyzer Setup
Start Frequency
50 MHz
Stop Frequency
500 MHz
Resolution Bandwidth
3 MHz
Video Bandwidth
3 kHz
Reference Level
20 dBm
Use your Spectrum Analyzer’s Peak Search function to find the desired reference signal.
The Analyzer should show the fundamental, and second and third harmonics. The
harmonics need to be adjusted so that the second harmonic is at -34 dBc and third
harmonic should typically be greater than or equal to -39 dBc as shown in Figure 6-29.
To adjust the harmonics, adjust R8, as shown in Figure 6-29 until the peaks of the
second and third harmonic are at the correct dB level. You may find that you can place
the second harmonic at -34 dBc but the third harmonic is less than -39 dBc. If this is the
case, continue adjusting R8 until the third harmonic is at –39dBc and the second
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......