5500A
Service Manual
6-102
6-130. Low Frequency Verification
This procedure provides an example of testing low frequency flatness using a 5.5 V
output. Follow the same procedure for testing other amplitudes, only compare results
against the flatness specification listed in Table 6-61.
1.
Program the Calibrator Mainframe for an output of 5.5 V @ 500 kHz. Press
O
on
the Calibrator Mainframe to activate the output.
2.
Allow the 5790A reading to stabilize. The 5790A should display approximately 1.94
V rms. Enter the 5790A reading in Column A of Table 6-61.
3.
Enter 50 kHz into the Calibrator Mainframe. Allow the 5790A reading to stabilize,
then enter the 5790A reading in Column B of Table 6-61.
4.
Enter the next frequency listed in Table 6-61. Allow the 5790A reading to stabilize,
then enter the reading into Column A of the table.
5.
Enter 50 kHz into the Calibrator Mainframe. Allow the 5790A reading to stabilize,
then enter the 5790A reading in Column B of Table 6-61.
6.
Repeat steps 4 and 5 for all of frequencies listed in Table 6-61. Continue until you
have completed Columns A and B.
7.
When you have completed Columns A and B, press
Y
to remove the Calibrator
Mainframe’s output. Complete Table 6-61 by performing the calculations for
column C. Compare Column C to the specifications listed in the final column.
Table 6-61. Low Frequency Flatness Verification at 5.5 V
Calibrator
Mainframe
Frequency
A
B
50 kHz
C
Calibrator Mainframe
Flatness Specification (%)
500 kHz
±
1.50 + 100
µ
V
1 MHz
±
1.50 + 100
µ
V
2 MHz
±
1.50 + 100
µ
V
5 MHz
±
1.50 + 100
µ
V
10 MHz
±
1.50 + 100
µ
V
Complete Columns A-C as follows:
A
Enter 5790A Reading (mV) for the present frequency.
B
Enter 5790A Reading (mV) for 50 kHz.
C
Compute and enter the Calibrator Mainframe Flatness Deviation (%): 100 * ((Column A entry)-
(Column B entry))/ (Column B entry)
6-131. High Frequency Verification
This procedure provides an example of testing high frequency flatness using a 5.5 V
output. Follow the same procedure for testing other amplitudes, only compare results
against the flatness specification listed in Table 6-62. For this voltage range, you will use
the model HP 8482A power sensor.
1.
Program the Calibrator Mainframe for an output of 5.5 V @ 30 MHz. Press
O
on
the Calibrator Mainframe to activate the output.
2.
Allow the power meter reading to stabilize. The power meter should display
approximately 75 mW. Enter the power meter’s reading in Column A of Table 6-62.
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......