5500A
Service Manual
6-80
4.
Ensure the HP 3458A reading is 0.0 V DC
±
100
µ
V.
5.
Press the GO ON blue softkey.
6.
Calibration voltages 33 V and greater will automatically put the Calibrator
Mainframe output in standby. When this occurs, press
O
on the Calibrator
Mainframe to activate the output. Allow the HP 3458A DC voltage reading to
stabilize. Enter the reading via the Calibrator Mainframe front panel keypad, then
press ENTER.
Note
The Calibrator Mainframe will warn when the entered value is out of
bounds. If this warning occurs recheck the setup and carefully re-enter the
reading insuring proper multiplier (i.e., m,
µ
, n, p). If the warning still
occurs, repair may be necessary.
7.
Repeat steps 6 until the Calibrator Mainframe display indicates that the next steps
calibrate ac voltage. Press the OPTIONS, then STORE CONSTS blue softkeys to
store the new calibration constants.
AC voltage must now be calibrated. Continue with the next section.
6-105. AC Square Wave Voltage Calibration
This procedure uses the same equipment and setup as DC Voltage calibration but
requires different settings on the HP 3458A. See “Calibration and Verification of Square
Wave Functions” earlier in this section for technical details on the procedure. DC
voltages are measured and entered in the Calibrator Mainframe to calibrate the AC
Voltage function.
Set up the Calibrator Mainframe to Cal ACV. Press OPTIONS and NEXT SECTION
blue softkeys until the display reads “The next steps calibrate -SC300 ACV”. Then
follow these steps to calibrate ac voltage:
1.
Press the GO ON blue softkey.
2.
Connect the Calibrator Mainframe’s SCOPE connector to the HP 3458A input, using
the BNC cable and the BNC(f) to Double Banana adapter.
3.
Set the HP 3458A to DCV, NPLC = .01, LEVEL 1, TRIG LEVEL, and the DELAY
to .0002 for measuring the upper part of the wave form (i.e. topline), and the
DELAY to .0007 for measuring the lower part of the wave form (i.e. baseline).
Manually range lock the HP 3458A to the range that gives the most resolution for
the topline measurements. Use this same range for the corresponding baseline
measurements at each step.
4.
For each calibration step, take samples for at least two seconds, using the HP 3458A
MATH functions to retrieve the average or mean value. See “Setup for Square Wave
Measurements” earlier in this chapter for more details.
The “true amplitude” of the wave form is the difference between the topline and
baseline measurements, correcting for the load resistance error. To make this
correction, multiply the readings by (0.5 * (50 + Rload)/Rload), where Rload =
actual feedthrough termination resistance if used.
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......