5500A
Service Manual
6-78
The Calibrator Mainframe first prompts the user to calibrate the DC Voltage function. If
another function is to be calibrated, alternately press the OPTIONS and NEXT
SECTION blue softkeys until the desired function is reached.
6-101. Calibration and Verification of Square Wave Functions
The AC Voltage and Edge functions have square wave voltages that need to be
calibrated and verified. The HP3458A digital multimeter can be programmed from either
the front panel or over the remote interface to make these measurements.
6-102. Overview of HP3458A Operation
The Hewlett-Packard 3458A digital multimeter is setup as a digitizer to measure the
peak-to-peak value of the signal. It is set to DCV, using various analog-to-digital
integration times and triggering commands to measure the topline and baseline of the
square wave signal.
6-103. Setup for Square Wave Measurements
By controlling the HP 3458A’s integration and sample time, it can be used to make
accurate, repeatable measurements of both the topline and baseline of the square wave
signals up to 10 kHz.
The HP 3458A is triggered by a change in input level. The trigger level is set to 1% of
the DCV range, with ac coupling of the trigger signal. The delay after the trigger event is
also changed for the of AC Voltage Square Wave and Edge functions. See Table 6-48
and Figure 6-19.
Table 6-48. AC Square Wave Voltage and Edge Settings for the HP3458A
Voltage
HP 3458A Settings
Input Frequency
NPLC
DELAY (topline)
DELAY (baseline)
10 Hz
1
.02 s
.07 s
100 Hz
.1
.002 s
.007 s
1 kHz
.01
.0002 s
.0007 s
5 kHz
.002
.00004 s
.00014 s
10 kHz
.001
.00002 s
.00007 s
Note
For this application, if making measurements of a signal > 1 kHz, the HP
3458A has been known to have .05% to .1% peaking in the 100 mV range.
For these signals, lock the HP 3458A to the 1 V range.
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......