5500A
Service Manual
6-60
Set the Calibrator Mainframe to SCOPE mode, with the Overload menu on the display.
Connect the BNC cable to the Calibrator Mainframe SCOPE connector. Then follow
these steps to verify the overload function.
1.
Connect the 50
Ω
feedthrough termination to the end of the BNC cable.
2.
Program the Calibrator Mainframe output for 5.000 V, DC (OUT VAL blue
softkey), and time limit = 60 s (T LIMIT blue softkey).
3.
Press
O
on the Calibrator Mainframe to activate the output and verify that the
OPR display timer increments.
4.
Remove the 50
Ω
feedthrough termination before 60 seconds and verify that
Calibrator Mainframe goes to STBY.
5.
Reconnect the 50
Ω
feedthrough termination to the end of the BNC cable.
6.
Program the Calibrator Mainframe output for 5.000 V, AC (OUT VAL blue
softkey).
7.
Press
O
on the Calibrator Mainframe to activate the output and verify that the
OPR display timer increments.
8.
Remove the 50
Ω
feedthrough termination before 60 seconds and verify that
Calibrator Mainframe goes to STBY.
6-74. SC600 Hardware Adjustments
Hardware adjustments must be made to the leveled sine and edge functions each time the
SC600 is repaired. In addition to the adjustment procedures, this section provides lists of
the required equipment and some recommendations on models that have the capabilities
required by these procedures. Equivalent models can be substituted if necessary.
6-75. Equipment
Required
The following equipment is necessary for performing the hardware adjustments
described in this section. The models listed are recommended for providing accurate
results.
•
Standard adjustment tool for adjusting the pots and trimmer caps
•
Extender Card
•
Oscilloscope Mainframe and Sampling Head (Tektronix 11801 with SD-22/26 or
Tektronix TDS 820 with 8 GHz bandwidth)
•
10 dB Attenuator (Weinschel 9-10 (SMA), or Weinschel 18W-10, or equivalent)
•
Cable provided with SC600
•
Spectrum Analyzer (Hewlett-Packard 8590A)
6-76. Adjusting the Leveled Sine Wave Function
There are two adjustment procedures that need to be made for the leveled sine wave
function. The first procedure adjusts the balance out of the LO VCO so that the signal is
balanced between the two VCOs. The second procedure adjusts the harmonics.
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......