5500A
Service Manual
6-44
6-60. Leveled Sine Wave Flatness Verification
Leveled Sine Wave flatness verification is divided into two frequency bands: 50 kHz to
10 MHz (low frequency) and >10 MHz to 600 MHz (high frequency). The equipment
setups are different for each band. Leveled Sine Wave flatness is measured relative to 50
kHz. This is determined directly in the low frequency band. The high frequency band
requires a “transfer” measurement be made at 10 MHz to calculate a flatness relative to
50 kHz.
6-61. Equipment Setup for Low Frequency Flatness
All low frequency flatness procedures use the following equipment.
•
5790A/03 AC Measurement Standard with Wideband option
•
BNC(f) to Type N(m) adapter
•
BNC cable supplied with the SC600
Connect the Calibrator Mainframe SCOPE connector to the 5790A WIDEBAND input
with the BNC(f) to Type N(m) adapter as shown in Figure 6-10. Set the 5790A to
AUTORANGE, digital filter mode to FAST, restart fine, and Hi Res on.
POWER
I
O
5790A
AC MEASUREMENT
STANDARD
1000V RMS MAX
1000V RMS MAX
3V RMS MAX
INPUT 1
INPUT 2
SHUNT
HI
LO
GUARD
GROUND
SHELL FLOATING
10V PEAK
MAX
WIDEBAND
7V RMS MAX
MAX
10V PK
SHELL FLOATING
INPUT1
SHUNT
INPUT1 INPUT1
INPUT1
2.2 mV
22 mV
220 mV
7 mV
70 mV
700 mV
2.2 V
22 V
220 mV
7V 70V
700V
1kV
2.2 mV
AUTO
MAN
VIEW
REF
UTIL
MENUS
SPEC
CLEAR
DELETE
ENTER
0
1
2
3
4
5
6
7
8
9
+/-
.
POWER
I
O
0
•
1
2
3
4
5
6
7
8
9
ENTER
M
k
m
V
Hz
FIELD
EDIT
/
+
F
OPR
EARTH
SCOPE
BOOST
MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
µ
n
p
W
dBm
sec
¡C
A
MULT
x
DIV
÷
OUT
TRIG
5500A CALIBRATOR
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL
AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
STBY
om034f.eps
Figure 6-10. Connecting the Calibrator Mainframe to the 5790A AC Measurement Standard
6-62. Equipment Setup for High Frequency Flatness
All high frequency flatness procedures use the following equipment.
•
Hewlett-Packard E4418A Power Meter
•
Hewlett-Packard 8482A and 8481D Power Sensors
•
BNC(f) to Type N(f) adapter
•
BNC cable supplied with the Calibrator Mainframe
Note
When high frequencies at voltages below 63 mV p-p are verified, use the
8481D Power Sensor. Otherwise, use the 8482A Power Sensor.
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......