SC600 Option
Calibration and Verification of Square Wave Voltage Functions
6
6-25
6-40. High Frequency Calibration
Connect the Calibrator Mainframe SCOPE connector to the power meter and power
sensor as described under “Equipment Setup for High Frequency Flatness”.
Follow these steps to calibrate high frequency Leveled Sine Wave flatness for the
amplitude being calibrated.
1.
Press the GO ON blue softkey.
2.
Establish the 10 MHz reference:
•
Press the power meter SHIFT key, then FREQ key and use the arrow keys to
enter the power sensor’s 10 MHz Cal Factor. Ensure that the factor is correct,
then press the power meter ENTER key.
•
Allow the power meter reading to stabilize.
•
Press the Power meter REL key.
3.
Press the GO ON blue softkey.
4.
Press the power meter SHIFT key, then FREQ key and use the arrow keys to enter
the power sensor’s Cal Factor for the frequency displayed on the Calibrator
Mainframe. Ensure that the factor is correct, then press the power meter ENTER
key.
5.
Adjust the amplitude using the Calibrator Mainframe front panel knob until the
power sensor reading matches the 10 MHz reference within 0.1%.
6.
Repeat steps 1 to 5 until the Calibrator Mainframe display indicates that either the
reference frequency is now 50 kHz or that the next steps calibrate pulse width.
Repeat the low frequency calibration procedure for the next amplitude unless the
Calibrator Mainframe display indicates that the next steps calibrate pulse width.
Press the OPTIONS, then STORE CONSTS blue softkeys to store the new
calibration constants.
6-41. Pulse Width Calibration
This procedure uses the following equipment:
•
High Frequency Digital Storage Oscilloscope: Tektronix 11801 with Tektronix SD-
22/26 sampling head
•
3 dB attenuator, 3.5 mm (m/f)
•
BNC(f) to 3.5 mm(m) adapter (2)
•
BNC cable supplied with the SC600
•
second BNC cable
Press the OPTIONS and NEXT SECTION blue softkeys until the display reads “Set up
to measure Pulse Width”. Then follow these steps to calibrate pulse width:
1.
Connect the BNC cable supplied with the SC600 to the Calibrator Mainframe’s
SCOPE connector. Connect the other end of the BNC cable to one BNC(f) to 3.5
mm(m) adapter then to the DSO’s sampling head through the 3 dB attenuator.
2.
Using the second BNC(f) to 3.5 mm(m) adapter and BNC cable, connect the
Calibrator Mainframe’s TRIG OUT connector to the 11801’s Trigger Input.
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......