SC600 Option
Calibration and Verification of Square Wave Voltage Functions
6
6-23
measurements at each step. Note that in the EDGE function, the topline is very near
0 V, and the baseline is a negative voltage.
3.
For each calibration step, take samples for at least two seconds, using the HP 3458A
MATH functions to enter the average or mean value. See “Setup for SC600 Edge
and Wave Generator Measurements” for more details.
The “true amplitude” of the wave form is the difference between the topline and baseline
measurements, correcting for the load resistance error. To make this correction, multiply
the readings by (0.5 * (50 + Rload)/Rload), where Rload = actual feedthrough
termination resistance.
6-37. Leveled Sine Wave Amplitude Calibration
This procedure uses the following equipment:
•
5790A AC Measurement Standard
•
BNC(f) to Double Banana Plug Adapter
•
50
Ω
feedthrough termination
•
BNC cable supplied with the SC600
Press the OPTIONS and NEXT SECTION blue softkeys until the display reads “Set up
to measure leveled sine amplitude”. Then follow these steps to calibrate Leveled Sine
Wave amplitude.
1.
Connect the BNC cable to the Calibrator Mainframe’s SCOPE connector. Connect
the other end of the BNC cable to the 50
Ω
feedthrough termination then to the
5790A INPUT 2 using the BNC(f) to Double Banana adapter.
2.
Set the 5790A to AUTORANGE, digital filter mode to FAST, restart fine, and Hi
Res on.
3.
Press the GO ON blue softkey.
4.
Press
O
to activate operating mode on the Calibrator Mainframe.
5.
Allow the 5790A rms reading to stabilize. Multiply the 5790A reading by (0.5 * (50
+ Rload) / Rload), where Rload = the actual feedthrough termination resistance, to
correct for the resistance error. Enter the corrected rms reading via the Calibrator
Mainframe front panel keypad, then press ENTER.
Note
The Calibrator Mainframe will warn when the entered value is out of
bounds. If this warning occurs recheck the setup and calculation and
carefully re-enter the corrected rms reading insuring proper multiplier
(i.e., m, u, n, p). If the warning still occurs, repair may be necessary.
6.
Repeat step 5 until the Calibrator Mainframe display indicates that the next steps
calibrate Leveled Sine flatness. Press the OPTIONS, then STORE CONSTS blue
softkeys to store the new calibration constants.
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......