Service Manual
6-22
Note
The Calibrator Mainframe will warn when the entered value is out of
bounds. If this warning occurs recheck the setup and carefully re-enter
the reading insuring proper multiplier (i.e., m, u, n, p). If the warning still
occurs, repair may be necessary.
3.
Repeat step 2 until the Calibrator Mainframe display indicates that the next steps
calibrate WAVEGEN. Press the OPTIONS, then STORE CONSTS blue softkeys
to store the new calibration constants.
6-35. Wave Generator Calibration
This procedure uses the following equipment:
•
Hewlett-Packard 3458A Digital Multimeter
•
BNC(f) to Double Banana adapter
•
BNC cable supplied with the SC600
Within the calibration menu, press the OPTIONS and NEXT SECTION blue softkeys
until the display reads “WAVEGEN Cal:”. Then follow these steps to calibrate the Wave
Generator:
1.
Connect the Calibrator Mainframe’s SCOPE connector to the HP 3458A input, using
the BNC cable and the BNC(f) to Double Banana adapter.
2.
Set the HP 3458A to DCV, NPLC = .01, LEVEL 1, TRIG LEVEL, and the DELAY
to .0002 for measuring the upper part of the wave form (i.e. topline), and the
DELAY to .0007 for measuring the lower part of the wave form (i.e. baseline).
Manually range lock the HP 3458A to the range that gives the most resolution for
the topline measurements. Use this same range for the corresponding baseline
measurements at each step.
3.
For each calibration step, take samples for at least two seconds, using the HP 3458A
MATH functions to retrieve the average or mean value. See “Setup for SC600 Edge
and Wave Generator Measurements” for more details.
6-36. Edge Amplitude Calibration
This procedure uses the following equipment:
•
Hewlett-Packard 3458A Digital Multimeter
•
BNC(f) to Double Banana adapter
•
BNC cable supplied with the SC600
•
50
Ω
feedthrough termination
Refer to Figure 6-3 for the proper setup connections. Press the OPTIONS and NEXT
SECTION blue softkeys until the display reads “Set up to measure fast edge
amplitude”. Then follow these steps to calibrate edge amplitude:.
1.
Connect the Calibrator Mainframe’s SCOPE connector to the HP 3458A input, using
the BNC cable and the BNC(f) to Double Banana.
2.
Set the HP 3458A to DCV, NPLC = .01, LEVEL 1, TRIG LEVEL, and the DELAY
to .0002 for measuring the upper part of the wave form (i.e. topline), and the
DELAY to .0007 for measuring the lower part of the wave form (i.e. baseline).
Manually lock the HP 3458A to the range that gives the most resolution for the
baseline measurements. Use this same range for the corresponding baseline
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......