![Fluke 522A/6 Operator'S Manual Download Page 44](http://html1.mh-extra.com/html/fluke/522a-6/522a-6_operators-manual_2310337044.webp)
5522A
Operators Manual
1-24
AC Voltage, DC Offset
Range
[1]
(Normal Channel)
Offset Range
[2]
Max Peak
Signal
1-Year Absolute Uncertainty,
tcal
±
5
°
C
[3]
±
(% dc floor)
Sine Waves (rms)
3.3 to 32.999 mV
0 to 50 mV
80 mV
0.1 + 33
μ
V
33 to 329.999 mV
0 to 500 mV
800 mV
0.1 + 330
μ
V
0.33 to 3.29999 V
0 to 5 V
8 V
0.1 + 3300
μ
V
3.3 to 32.9999 V
0 to 50 V
55 V
0.1 + 33 mV
Triangle Waves and Truncated Sine Waves (p-p)
9.3 to 92.999 mV
0 to 50 mV
80 mV
0.1 + 93
μ
V
93 to 929.999 mV
0 to 500 mV
800 mV
0.1 + 930
μ
V
0.93 to 9.29999 V
0 to 5 V
8 V
0.1 + 9300
μ
V
9.3 to 93.0000 V
0 to 50 V
55 V
0.1 + 93 mV
Square Waves (p-p)
6.6 to 65.999 mV
0 to 50 mV
80 mV
0.1 + 66
μ
V
66 to 659.999 mV
0 to 500 mV
800 mV
0.1 + 660
μ
V
0.66 to 6.59999 V
0 to 5 V
8 V
0.1 + 6600
μ
V
6.6 to 66.0000 V
0 to 50 V
55 V
0.1 + 66 mV
[1]
Offsets are not allowed on ranges above the highest range shown above.
[2]
The maximum offset value is determined by the difference between the peak value of the selected voltage output and the
allowable maximum peak signal. For example, a 10 V p-p square wave output has a peak value of 5 V, allowing a maximum
offset up to
±
50 V to not exceed the 55 V maximum peak signal. The maximum offset values shown above are for the
minimum outputs in each range.
[3]
For frequencies 0.01 to 10 Hz, and 500 kHz to 2 MHz, the offset uncertainty is 5 % of output,
±
1 % of the offset range.
AC Voltage, Square Wave Characteristics
Risetime @
1 kHz
Typical
Settling Time @
1 kHz Typical
Overshoot
@ 1 kHz
Typical
Duty Cycle Range
Duty Cycle Uncertainty
<1
μ
s
<10
μ
S to 1 % of
final value
<2 %
1 % to 99 % <3.3 V p-p.
0,01 Hz to 100 kHz
±
(0.02 % of 100 ns), 50 % duty cycle
±
(0.05 % of 100 ns), other duty cycles
from 10 % to 90 %
±
(0.8 % of 100 ns)
AC Voltage, Triangle Wave Characteristics
Linearity to 1 kHz
Aberrations
0.3 % of p-p value, from 10 % to 90 % point
<1 % of p-p value, with amplitude >50 % of range
AC Current (Non-Sine Wave)
Triangle Wave &
Truncated Sine Wave
Range
p-p
Frequency
1-Year Absolute Uncertainty tcal
±
5
°
C
±
(% of % of range)
Max Current
Resolution
10 to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
0.047 to
0.92999 mA
[1]
1 to 10 kHz
10 + 2
Six digits
10 to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
0.93 to
9.29999 mA
[1]
1 to 10 kHz
10 + 2
Six digits
10 to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
9.3 to
92.9999 mA
[1]
1 to 10 kHz
10 + 2
Six digits
Summary of Contents for 522A/6
Page 4: ......
Page 14: ...5522A Operators Manual x...
Page 18: ...5522A Operators Manual xiv...
Page 22: ...5522A Operators Manual 1 2...
Page 46: ...5522A Operators Manual 1 26...
Page 48: ...5522A Operators Manual 2 2...
Page 54: ...5522A Operators Manual 2 8...
Page 56: ...5522A Operators Manual 3 2...
Page 190: ...5522A Operators Manual 6 2...
Page 240: ...5522A Operators Manual 7 2...
Page 260: ...5522A Operators Manual 8 2...
Page 314: ...5522A Operators Manual 9 52...
Page 370: ...5522A Operators Manual 10 56...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 414: ...5522A Operators Manual B 2...
Page 416: ...5522A Operators Manual B 4...