Figure 10-2. Tunnel Diode Pulser Connections
The Leveled Sine Wave Function
The Leveled Sine Wave (LEVSINE) function uses a leveled sine wave, whose amplitude
remains relatively constant over a range of frequencies, to check the oscilloscope’s
bandwidth. When you check your oscilloscope, you change the wave’s frequency until
the amplitude displayed on the oscilloscope drops 30 %, which is the amplitude that
corresponds to the -3 dB point. Default values are 30 mV p-p, 50 kHz.
To access the LEVSINE menu, press the softkey under MODE until “levsine” appears.
gjh056.eps
You can press the MODE softkey to cycle through the functions in the order shown, or
you can press
P
to return directly to the SCOPE menu.
Each option in the LEVSINE menu is described below:
Summary of Contents for 522A/6
Page 4: ......
Page 14: ...5522A Operators Manual x...
Page 18: ...5522A Operators Manual xiv...
Page 22: ...5522A Operators Manual 1 2...
Page 46: ...5522A Operators Manual 1 26...
Page 48: ...5522A Operators Manual 2 2...
Page 54: ...5522A Operators Manual 2 8...
Page 56: ...5522A Operators Manual 3 2...
Page 190: ...5522A Operators Manual 6 2...
Page 240: ...5522A Operators Manual 7 2...
Page 260: ...5522A Operators Manual 8 2...
Page 314: ...5522A Operators Manual 9 52...
Page 370: ...5522A Operators Manual 10 56...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 414: ...5522A Operators Manual B 2...
Page 416: ...5522A Operators Manual B 4...