![Fluke 522A/6 Operator'S Manual Download Page 256](http://html1.mh-extra.com/html/fluke/522a-6/522a-6_operators-manual_2310337256.webp)
5522A
Operators Manual
7-18
Table 7-13. Verification Tests for Phase Accuracy, V and V
Range,
Normal
Output, V
Output,
Normal V
Frequency
Range,
AUX
Output
Output,
AUX
Phase
°
Lower
Limit
°
Upper
Limit
°
65 Hz
-0.10
0.10
400 Hz
-0.25
0.25
1 kHz
-0.50
0.50
5 kHz
-2.50
2.50
10 kHz
-5.00
5.00
30 kHz
0
-10.00 10.00
65 Hz
59.90
60.10
400 Hz
59.75
60.25
1 kHz
59.50
60.50
5 kHz
57.50
62.50
10 kHz
55.00
65.00
30 kHz
60
50.00 70.00
65 Hz
89.90
90.10
400 Hz
89.75
90.25
1 kHz
89.50
90.50
5 kHz
87.50
92.50
10 kHz
85.00
95.00
3.29999 3.00000
30 kHz
80.00
100.00
32.9999 30.0000 65
Hz
89.90 90.10
329.999 50.000 65
Hz
3.29999 V
3.00000 V
90
89.80 90.10
Summary of Contents for 522A/6
Page 4: ......
Page 14: ...5522A Operators Manual x...
Page 18: ...5522A Operators Manual xiv...
Page 22: ...5522A Operators Manual 1 2...
Page 46: ...5522A Operators Manual 1 26...
Page 48: ...5522A Operators Manual 2 2...
Page 54: ...5522A Operators Manual 2 8...
Page 56: ...5522A Operators Manual 3 2...
Page 190: ...5522A Operators Manual 6 2...
Page 240: ...5522A Operators Manual 7 2...
Page 260: ...5522A Operators Manual 8 2...
Page 314: ...5522A Operators Manual 9 52...
Page 370: ...5522A Operators Manual 10 56...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 414: ...5522A Operators Manual B 2...
Page 416: ...5522A Operators Manual B 4...