![Fluke 522A/6 Operator'S Manual Download Page 176](http://html1.mh-extra.com/html/fluke/522a-6/522a-6_operators-manual_2310337176.webp)
5522A
Operators Manual
5-34
Table 5-10. Response Data Types (cont.)
Data Type
Description
Binary Block Data
A special data type defined by the IEEE-488.2 standard. This type is
used in *PUD? query. It is defined as follows:
#(non-zero digit) (digits) (user data)
The non-zero digit specifies the number of characters that will follow in
the <digits> field. Characters allowed in the digits field are 0 through 9
(ASCII 48 through 57 decimal). The value of the number in the <digits>
field in decimal defines the number of user data bytes that follow in the
<user data> field. The maximum response is 64 characters.
Example: *
PUD “test1”; *PUD?
returns:
#205test1
Checking 5522A Status
The programmer has access to status registers, enable registers, and queues in the
Calibrator to indicate various conditions in the instrument as shown in Figure 5-8. Some
registers and queues are defined by the IEEE-488.2 standard. The rest are specific to the
Calibrator. In addition to the status registers, the Service Request (SRQ) control line, and
a 16-element buffer called the Error Queue provide status information. Table 5-11 lists
the status registers and gives the read/write commands and associated mask registers.
Table 5-11. Status Register Summary
Status Register
Read
Command
Write
Command
Serial Poll Status Byte (STB)
*STB?
⎯
Service Request Enable Register (SRE)
*SRE?
*SRE
Event Status Register (ESR)
*ESR?
⎯
Event Status Enable Register (ESE)
*ESE?
*ESE
Instrument Status Register (ISR)
ISR?
⎯
Instrument Status Change Register (ISCR)
ISCR?
⎯
ISCR 1 to 0 transition
ISCR0?
⎯
ISCR 0 to 1 transition
ISCR1?
⎯
Instrument Status Change Enable Register (ISCE)
ISCE?
ISCE
ISCE 1 to 0 transition
ISCE0?
ISCE0
ISCE 0 to 1 transition
ISCE1?
ISCE1
Each status register and queue has a summary bit in the Serial Poll Status Byte. Enable
registers are used to mask various bits in the status registers and generate summary bits in
the Serial Poll Status Byte. For IEEE-488 interface operation, the Service Request Enable
Register is used to assert the SRQ control line on detection of any status condition or
conditions the programmer chooses. For RS-232 interface operation, the SRQSTR string
is sent over the serial interface when the SRQ line is set. (See the
SRQSTR
command
description in Chapter 6 for more information.)
Summary of Contents for 522A/6
Page 4: ......
Page 14: ...5522A Operators Manual x...
Page 18: ...5522A Operators Manual xiv...
Page 22: ...5522A Operators Manual 1 2...
Page 46: ...5522A Operators Manual 1 26...
Page 48: ...5522A Operators Manual 2 2...
Page 54: ...5522A Operators Manual 2 8...
Page 56: ...5522A Operators Manual 3 2...
Page 190: ...5522A Operators Manual 6 2...
Page 240: ...5522A Operators Manual 7 2...
Page 260: ...5522A Operators Manual 8 2...
Page 314: ...5522A Operators Manual 9 52...
Page 370: ...5522A Operators Manual 10 56...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 414: ...5522A Operators Manual B 2...
Page 416: ...5522A Operators Manual B 4...