Front Panel Operation
Sample Applications
4
4-61
Warning
Ensure that the calibrator is in standby mode before making
any connection between the calibrator and Tester. Dangerous
voltages may be present on the leads and connectors.
Table 4-3. Watts Performance, Text Screen
Calibrator Outputs
Performance Limits
Normal
V ac @
Phase
in
AUX
mV ac
W/KW VA/KVA
VAR/KVAR
Model 41
Only
Phase
Harmonics
Screen
60 Hz
DEG.
@ 60
Hz
MIN MAX MIN MAX MIN MAX MIN MAX
5.0 V
0.0
30.0 mV
145
156
145
156
0
4
-2
2
8.0 V
0.0
30.0 mV
234
246
234
246
0
4
-2
2
100.0 V
157.0
150.0 mV
-14.3k -13.3k 14.5k 15.6k 5.4k 6.3k 155 159
100.0 V
157.0
360.0 mV
-37k
-29k 32k 40k 10k 18k 155 159
10.0 V
46.0
1.40 V
9.2
10.2 13.5
14.5 9.6 10.6
44 48
100.0 V
46.0
1.40 V
92
102
135
145
96
106
44
48
1.
Connect the calibrator to the Model 41 as shown in Figure 4-21.
Note
Voltage is connected to the Model 41 amps channel to simulate current
clamp operation (1 mV = 1 A).
5522A
CALIBRATOR
POWER HARMONICS
ANALYZER
41
®
gjh028.eps
Figure 4-21. Cable Connections for Testing a 40 Series Watts Function
Summary of Contents for 522A/6
Page 4: ......
Page 14: ...5522A Operators Manual x...
Page 18: ...5522A Operators Manual xiv...
Page 22: ...5522A Operators Manual 1 2...
Page 46: ...5522A Operators Manual 1 26...
Page 48: ...5522A Operators Manual 2 2...
Page 54: ...5522A Operators Manual 2 8...
Page 56: ...5522A Operators Manual 3 2...
Page 190: ...5522A Operators Manual 6 2...
Page 240: ...5522A Operators Manual 7 2...
Page 260: ...5522A Operators Manual 8 2...
Page 314: ...5522A Operators Manual 9 52...
Page 370: ...5522A Operators Manual 10 56...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 414: ...5522A Operators Manual B 2...
Page 416: ...5522A Operators Manual B 4...