Fluke 3125 User Manual Download Page 5

Symbol

Description

Canadian Standards Association

OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re-
fers to the level of Impulse Withstand Voltage protection provided. Equipment of
OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from
the fixed installation. Examples include household, office, and laboratory appliances.

C-TIC Australian EMC Mark

The European Waste Electrical and Electronic Equipment (WEEE) Directive
(2002/96/EC) mark.

1.2

Safety Information

Use this instrument only as specified in this manual. Otherwise, the protection
provided by the instrument may be impaired.

The following definitions apply to the terms “Warning” and “Caution”.

“Warning” identifies conditions and actions that may pose hazards to the
user.

“Caution” identifies conditions and actions that may damage the instru-
ment being used.

1.2.1

Warnings

To avoid personal injury, follow these guidelines.

BURN HAZARD – DO NOT

touch the plate or surrounding areas of the

unit.The temperature of the plate surface is the same as the actual temper-
ature shown on the display. If the unit is set at 400°C and the display
reads 400°C, the target surface is 400°C.The top sheet metal of the instru-
ment may exhibit extreme temperatures.

DO NOT

turn off the unit at

temperatures higher than 100°C. This could create a hazardous situation.
Select a set-point less than 100°C and allow the unit to cool before turn-
ing it off.

DO NOT

operate this unit without a properly grounded, properly polar-

ized power cord.

DO NOT

connect this unit to any other instrument or outlet other than the

2200 controller.

DO NOT

connect the 2200 to a non-grounded, non-polarized outlet.

The 2200 Controller and the 3125 Detachable Hot Plate are a matched

set.

DO NOT

mix and match sets.

HIGH VOLTAGE

is used in the operation of this equipment.

SEVERE

INJURY OR DEATH

may result if personnel fail to observe safety pre-

3125 Detachable Hotplate

User’s Guide

2

Summary of Contents for 3125

Page 1: ...3125 Detachable Hot Plate Users Manual Rev 622101 Hart Scientific ...

Page 2: ...e warranty period To obtain warranty service contact your nearest Hart authorized service center or send the product with a description of the difficulty postage and insurance prepaid FOB Destination to the nearest Hart au thorized service center Hart assumes no risk for damage in transit Following warranty repair the prod uct will be returned to Buyer transportation prepaid FOB Destination If Har...

Page 3: ...s 4 1 3 Authorized Service Centers 4 2 Introduction 7 3 Specifications and Environmental Conditions 9 3 1 Specifications 9 3 2 Environmental Conditions 9 4 Troubleshooting 11 4 1 Troubleshooting Problems Possible Causes and Solutions 11 4 2 Comments 12 4 2 1 EMC Directive 12 4 2 2 Low Voltage Directive Safety 12 i ...

Page 4: ...the instrument or in this manual Symbol Description AC Alternating Current AC DC Battery CE Complies with European Union Directives DC Double Insulated Electric Shock Fuse PE Ground Hot Surface Burn Hazard Read the User s Manual Important Information Off On 1 1 Before You Start Symbols Used Table 1 International Electrical Symbols ...

Page 5: ...To avoid personal injury follow these guidelines BURN HAZARD DO NOT touch the plate or surrounding areas of the unit The temperature of the plate surface is the same as the actual temper ature shown on the display If the unit is set at 400 C and the display reads 400 C the target surface is 400 C The top sheet metal of the instru ment may exhibit extreme temperatures DO NOT turn off the unit at te...

Page 6: ...r has been in a wet environment take necessary measures to remove moisture prior to apply ing power such as storage in a low humidity temperature chamber operating at 50 C for 4 hours or more The instrument generates extreme temperatures Precautions must be taken to prevent personal injury or damage to objects Sensors may be ex tremely hot when removed from the instrument Cautiously handle sen sor...

Page 7: ...nto the instrument causing damage The instrument is a precision instrument Although is has been designed for optimum durability and trouble free operation it must be handled with care Always carry the unit in an upright position The instrument should not be operated in excessively wet oily dusty or dirty environments Do not operate near flammable materials If a main supply power fluctuation occurs...

Page 8: ... 512 3437 E mail xingye han fluke com cn Fluke South East Asia Pte Ltd Fluke ASEAN Regional Office Service Center 60 Alexandra Terrace 03 16 The Comtech Lobby D 118502 SINGAPORE Phone 65 6799 5588 Telefax 65 6799 5588 E mail antng singa fluke com When contacting these Service Centers for support please have the following information available Model Number Serial Number Voltage Complete description...

Page 9: ...ection devices insure user and instrument safety and protection The Detachable Hot Plate see Figure 1 consists of a controlled plate made of aluminum A heater is attached to the bottom with an RTD temperature sensor used to control it The housing consists of a top cover and a base which serve as a clamp and strain relief point for the wiring The Reference Well is avail able for use with a 3 16 inc...

Page 10: ...olid state controlled Surface Plate 6061 aluminum top surface machine finished to 0 000032 0 0008 mm 3 8 96 mm diameter accessible Power 115 VAC 10 2 8 A or 230 VAC 10 1 4 A specify 50 60 Hz 325 W Weight 7 lb 3 2 kg with 2200 Controller NIST Traceable Calibration Data at 50 C 120 C 190 C 260 C 330 C and 400 C Safety OVERVOLTAGE Installation CATEGORY II Pollution Degree 2 per IEC1010 1 3 2 Environm...

Page 11: ...for temperature 31 C de creasing linearly to 50 at 40 C pressure 75kPa 106kPa mains voltage within 10 of nominal vibrations in the calibration environment should be minimized altitude does not effect the performance or safety of the unit 10 ...

Page 12: ...Reset Sequence Factory Reset Sequence Hold the SET and EXIT buttons down at the same time while powering up the instrument After the instrument displays init release the buttons The display shows init then displays 9132 and then displays the firmware version After performing the master reset sequence all of the configuration parameters are reset to their default values Reprogram the R0 ALPHA and D...

Page 13: ... certain point Incorrect High Limit parameter The High Limit parameter may be set below 400 C Check this value as described in the 2200 User s Guide 4 2 Comments 4 2 1 EMC Directive Hart Scientifics equipment has been tested to meet the European Electromag netic Compatibility Directive EMC Directive 89 336 EEC The Declaration of Conformity for your instrument lists the specific standards to which ...

Reviews: