1587/1577
Calibration
26
Table 6. Calibration Adjustment Steps
Switch Position
(Function)
Input
Terminal
Calibration Adjustment Step
Input Value
C-01
600.0 mV, 60 Hz
C-02
(Model 1587 only)
[1]
600.0 mV, 5 kHz
C-03
6.000 V, 60 Hz
C-04
(Model 1587 only)
[1]
6.000
V,
5
kHz
C-05
60.00 V, 60 Hz
C-06
(Model 1587 only)
[1]
60.0 V, 5 kHz
C-07
600.0 V, 60 Hz
AC Volts
V/COM
C-08
(Model 1587 only)
[1]
600.0
V,
2
kHz
C-09
6.000 V, 0 Hz
C-10
60.00 V, 0 Hz
DC Volts
C-11
600.0 V, 0 Hz
C-12
600.0 mV, 0 Hz
DC Millivolts
C-13
60.00 mV, 0 Hz
C-14
600.0
Ω
C-15
6.000 k
Ω
C-16
60.00 k
Ω
C-17
600.0 k
Ω
C-18
6.000 M
Ω
C-19
0.000
Ω
Ohms
C-20
50.0 M
Ω
Continuity
C-21
600.0
Ω
Diode Test
C-22
(Model 1587 only)
[1]
6.000
V,
0
Hz
C-23
60.00 mA, 60 Hz
C-24
400.0 mA, 60 Hz
C-25
60.00 mA, 0 Hz
milliamps mA/COM
C-26
400.0 mA, 0 Hz
Insulation mA/COM
C-27
1.0 mA, 0 Hz
XW
Warning
To avoid electrical shock hazard, remove all test leads from
the UUT before performing the following step.
Insulation None
[2]
C-28
None (press BLUE to start
internal cal)
Notes:
1.
Calibration step will be skipped on Model 1577.
2.
Internal HV output calibration. Ensure no lead is inserted in Insulation (+) jack.