1587/1577
Calibration
16
Performance Test
The following series of tests comprise a performance test for verifying the accuracy
of the Meter (UUT) and its performance level. The performance test is recommended
as an acceptance test for incoming inspection and as a calibration procedure for
periodically ensuring the accuracy of the Meter. Fluke recommends running the
performance test at least once a year.
No adjustments are required during the performance test, and it is not necessary to
open the case. If the Meter does not pass all parts of the performance tests, repair
and/or calibration adjustment are required. A calibration adjustment procedure is
given later in this manual. If significant repairs are required, contact Fluke as
described toward the front of this manual. If user repairs are appropriate, refer to the
list of user-replaceable parts toward the end of this manual
Testing the 1577 and 1587 DMM Functions
Use the following procedure to verify the accuracy of the measurement functions.
1.
Connect the Calibrator to the
J
and
COM
input terminals on the UUT.
2.
Turn the rotary switch to the function listed in each step of Table 2.
3.
Apply the input level for each step listed in Table 2.
4.
Compare the reading on the UUT display with the Display Reading in Table 2.
If the display reading falls outside of the range shown in Table 2, the UUT does
not meet specification.
Table 3. DMM Accuracy Tests
UUT Display Reading
Step
UUT
Function
UUT
Range
UUT
Input Level
Fluke 1577
Fluke 1587
1
mV ac
600.0 mV
30 mV, 60Hz
29.4 to 30.6
29.4 to 30.6
2
V ac
6.000V
3V, 60Hz
2.937 to 3.063
2.967 to 3.033
3
V ac
6.000V
3V, 400Hz
n/a
2.937 to 3.063
Push
G
(blue) for 1 second for low-pass filter band-pass test
4
V ac
6.000V
3V, 400Hz
n/a
2.817 to 3.063
5
V ac
6.000V
3V, 800Hz
n/a
1.680 to 2.670
Push
G
(blue) for 1 second to remove low-pass filter
6
V ac
60.00V
60V, 50Hz
58.77 to 61.23
59.37 to 60.63
7
V ac
600.0V
600V, 50Hz
587.7 to 612.3
593.7 to 606.3
8
V ac
1000V
1000V, 50 Hz
977 to 1023
n/a
9
V ac
1000V
1000V, 1 kHz,
n/a
977 to 1023
10
Hz, V ac
99.99 Hz
*10 Hz, 0.15V
n/a
9.98 to 10.02
11
Hz, V ac
99.99 kHz
*95kHz, 0.15V
n/a
94.89 to 95.11
12
Hz, V ac
99.99 kHz
*90 kHz, 2V
n/a
89.90 to 90.10
13
V dc
6.000V
0.3V
0.298 to 0.302
0.298 to 0.302
14
V dc
6.000V
6V
5.986 to 6.014
5.993 to 6.007
15
V dc
60.00V
60V
59.86 to 60.14
59.93 to 60.07