The DELAY parameter (0 to 32,767) sets the minimum delay time, in seconds,
between each measurement. Use the numeric buttons to enter a value and press
ENTER
. Measurements may take longer than the specified delay time, if nec-
essary. This value also affects the time axis of the graph window.
The SEQ TIMER parameter (0 to 10,000) times the start of each scan sequence.
For instance, if SEQ TIMER is set to 300, the scan sequence will run once ev-
ery five minutes. If the SEQ timer value is 0, scanning will run continuously.
This applies only with the SCAN CHAN and SCAN/PRIM scan modes. The
SEQ TIMER value is ignored when the time scale of the graph display is calcu-
lated. Consider disabling the average function (Section 5.1.5) when using long
measurement intervals.
5.1.2
Primary Channel
The
PRIM CHAN
soft-key selects the primary input channel. The channel is
selected from a window showing a list of all available input channels. Each
channel in the list is identified with the device name, the channel number, and
probe serial number. The channel is selected using the
UD
buttons and press-
ing
ENTER
.
SELECT THE PRIMARY CHANNEL
>SPRT 01, PROBE 566-011
SPRT 02, PROBE 566-012
TCS
03, PROBE 1341
TCS
04, PROBE 1342
TCS
05, PROBE 1343
TCS
06, PROBE 1344
Selecting a primary channel will also set the SCAN MODE to PRIM CHAN
(see Section 5.1.4).
5.1.3
Scan Channels
The
SCAN CHAN
soft-key allows you to select channels to scan with the
SCAN and SCAN/PRIM measurement modes. A window shows a list of all
channels and the ON/OFF state of each. Only channels that are set to ON will
be measured when the SCAN MODE is SCAN or SCAN/PRIM. Each channel
in the list is identified with the device name and the channel number. You can
scroll through the list using the
UD
buttons. The indicated channel can be
toggled on or off using the
LR
buttons. It is not necessary to press
ENTER
.
Press
EXIT
to exit.
31
5 Soft-Key Functions
Input Menu