Preparation
PCU Priming
2
2-5
Table 2-1. Test Port Insert - Parts List
Item Description Part
1 O
Ring
3865142
2
Test Port Insert
3919892
3 O
Ring
3883397
To calibrate panel-mounted gauges with pressure connections in the rear, use an
Angle Adapter P5543, (see Chapter 8, Ancillary Equipment).
PCU Priming
If the instrument is to be primed for the first time after drainage for storage/transportation,
the following points apply:
1.
Remove PCU assembly — see Chapter 5, PCU Assembly.
2.
Remove filler plug.
3.
Using the filler bottle, fill the chamber through the top port until the fluid level
reaches the PCU O-ring. Allow time for the fluid level to stabilize.
4.
Continue this procedure until the fluid level reaches the bottom of the filler plug
port — DO NOT OVERFILL. If this occurs, wipe up any spills, and allow any
excess fluid to drain out of the port.
5.
Replace the filler plug.
6.
Replace the PCU assembly.
1
2
gmg08.eps
Figure 2-8. Sectional View - PCU Housing