Multifunction Electrical Tester Calibrator
Verification
73
e. Use the numerical keypad, cursor keys, or rotary knob to set the time interval on
Product display in ms.
Figure 40. Trip Time Verification Screen
f. If the counter does not indicate 0 ms, reset the counter display.
g. Push
O
on the Product. The Product generates passive square wave with a
frequency of 2/T.
h. Compare the counter with the Product preset time value with the limits in
.
METER
Follow these steps to verify the METER functions.
1. DC/AC multimeter voltage test:
a. Connect the 5522A voltage output terminals to V and COM terminals of the Product.
b. Set the frequency and output voltage as in
c. Select the
METER
function, and the
Multimeter
mode on the Product.
d. Select the appropriate
AC
or
DC
mode.
e. Switch the 5522A output to ON.
f. Compare the measurement on the Product with the limits in
2. DC/AC multimeter current test:
a. Connect the 5522A current output terminals to the A and COM terminals on the
Product. Set the frequency and output voltage as in
b. Select the
METER
function, and the
Multimeter
mode on the Product.
c. Select the appropriate
AC
or
DC
mode.
d. Switch the 5522A output to ON.
e. Compare the measurement on the Product with the limits in
.